DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2016-44278
- Event Type
- Malfunction
- Date Received
- June 7, 2016
- Date of Event
- March 25, 2016
- Report Date
- May 11, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- UDI-DI
- 00386270000064
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
STUDY COORDINATOR CONTACTED DEXCOM ON (B)(6) 2016 TO REPORT THAT ON (B)(6) 2016, PATIENT EXPERIENCED AN INTERMITTENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
THE COMPLAINT RECEIVER DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THERE WAS NO FAILURE DETECTED RELATED TO THE COMPLAINT. A PAIRING TEST WAS PERFORMED WITH A KNOWN GOOD TRANSMITTER AND PASSED. A REVIEW OF THE DOWNLOADED RECEIVER LOG DID NOT CONFIRM THE REPORTED EVENT OF AN INTERMITTENT OUT OF RANGE. A ROOT CAUSE COULD NOT BE DETERMINED. ADDITIONALLY, A TRANSMITTER DEVICE WAS RETURNED FOR EVALUATION (SERIAL #(B)(4), LOT #5192811). THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. VOLTAGE TESTING WAS PERFORMED AND THE TEST FAILED. THE REPORTED EVENT OF INTERMITTENT ANTENNA ICON WAS NOT CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 357873 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649 | 140514406 | 00386270000064 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |