PREMILENE 1 (4) 75CM HRT37 (M)
Report
- Report Number
- 2916714-2016-00426
- Event Type
- Malfunction
- Date Received
- June 7, 2016
- Report Date
- June 7, 2016
- Manufacturer
- B.BRAUN SURGICAL SA
- Product Code
- GAW
- PMA / PMN Number
- K980703
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
(B)(4) (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER). EXEMPTION NUMBER: E2014012. MANUFACTURING SITE EVALUATION: SAMPLES RECEIVED: 16 UNOPENED RACEPACKS AND 1 OPENED. ANALYSIS AND RESULTS: THERE IS ONE PREVIOUS COMPLAINT OF THIS CODE BATCH. MANUFACTURED (B)(4) UNITS OF THIS CODE BATCH, THERE ARE NO UNITS IN STOCK. RECEIVED 16 CLOSED SAMPLES AND 1 OPEN SAMPLE WITH THE NEEDLE DETACHED FROM THE THREAD, AND THE THREAD IS STILL WOUND ON THE PACK. TESTED THE NEEDLE ATTACHMENT OF THE CLOSED SAMPLES RECEIVED AND THE RESULTS DO NOT FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILLED THE OEM REQUIREMENTS. FINAL CONCLUSION: COMPLAINT JUSTIFIED: TAKING INTO ACCOUNT THAT THE RESULTS OF CLOSED SAMPLES RECEIVED DO NOT FULFILL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP). WE APOLOGISE FOR ANY INCONVENIENCE THAT THIS ISSUE MAY HAVE CAUSED AND THANK YOU FOR YOUR COLLABORATION. ACTIONS ON DISTRIBUTED PRODUCT OF THIS REFERENCE/BATCH: REPLACE THIS CODE/BATCH TO THE CUSTOMER OR A REFUND. CORRECTIVE/PREVENTIVE ACTIONS: THIS COMPLAINT WILL BE INCLUDED IN THE ANALYSIS OF TRENDING, AND CORRECTIVE ACTION WILL BE OPEN IF APPLIES.
COUNTRY OF COMPLAINT: SPAIN. THE NEEDLES ARE DETACHED FROM THE SUTURES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 358645 | PREMILENE 1 (4) 75CM HRT37 (M) | SUTURES | GAW | B.BRAUN SURGICAL SA | C2090463 | 114193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |