TRUERESULT
Report
- Report Number
- 1052693-2016-01047
- Event Type
- Malfunction
- Date Received
- June 2, 2016
- Date of Event
- May 10, 2016
- Report Date
- June 2, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. RETURNED TEST STRIPS EVALUATED WITH DEFECT FOUND. THE MOST LIKELY ROOT CAUSE OF PQQ-PD LOT PT2717 READ LOW IN METER GLUCOSE READINGS IS DUE TO RETURNED STRIPS WERE LEFT OUTSIDE OF RETURNED VIAL FOR PROLONGED OF TIME OR WERE EXPOSED TO A HUMID ENVIRONMENT.
CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. CUSTOMER STATED HIS NORMAL BLOOD SUGAR RANGE OF READING IS BETWEEN 80-120 MG/DL FASTING AND 120-140 MG/DL WITHIN TWO HOURS AFTER EATING. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL INTERVENTION IS NOT REPORTED AT THIS TIME. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 132 MG/DL AND 128 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/28/2019 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350439 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUERESULT | PT2717 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |