FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 5694952 · Received June 2, 2016

Report

Report Number
1052693-2016-01047
Event Type
Malfunction
Date Received
June 2, 2016
Date of Event
May 10, 2016
Report Date
June 2, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. RETURNED TEST STRIPS EVALUATED WITH DEFECT FOUND. THE MOST LIKELY ROOT CAUSE OF PQQ-PD LOT PT2717 READ LOW IN METER GLUCOSE READINGS IS DUE TO RETURNED STRIPS WERE LEFT OUTSIDE OF RETURNED VIAL FOR PROLONGED OF TIME OR WERE EXPOSED TO A HUMID ENVIRONMENT.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. CUSTOMER STATED HIS NORMAL BLOOD SUGAR RANGE OF READING IS BETWEEN 80-120 MG/DL FASTING AND 120-140 MG/DL WITHIN TWO HOURS AFTER EATING. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL INTERVENTION IS NOT REPORTED AT THIS TIME. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 132 MG/DL AND 128 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/28/2019 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350439 TRUERESULT BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUERESULT PT2717

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY