ELEVATOR #301
Report
- Report Number
- 0001032347-2016-00265
- Event Type
- Malfunction
- Date Received
- June 2, 2016
- Report Date
- April 25, 2016
- Manufacturer
- BIOMET MICROFIXATION
- Product Code
- EMJ
- PMA / PMN Number
- PEXEMPT
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE WARNINGS IN THE PACKAGE INSERT STATE THIS TYPE OF EVENT CAN OCCUR. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. A FOLLOW UP REPORT WILL BE SENT UPON COMPLETION OF THE DEVICE EVALUATION. THIS IS REPORT 3 OF 5 FOR THE SAME EVENT. REPORTS 1,2, 4, AND 5 ARE REPORTED ON MFR: 0001032347-2016-00263, 0001032347-2016-00264, 0001032347-2016-00266, 0001032347-2016-00267.
THE PRODUCT WAS RETURNED FOR EVALUATION. ACCORDING TO THE PRODUCT EVALUATION, THE COMPLAINT IS CONFIRMED AS THE TIP OF THE ELEVATOR IS SLIGHTLY BENT. THE MOST-LIKELY, UNDERLYING CAUSE OF THE COMPLAINT WAS DETERMINED TO BE EXCESSIVE FORCE. THE INSTRUCTIONS FOR USE STATES, "THE TIP OF THE INSTRUMENT IS EXTREMELY THIN AND DELICATE; CARE SHOULD BE TAKEN TO AVOID APPLYING SIGNIFICANT PRESSURE TO THE TIP." THE NON-CONFORMANCE DATABASE WAS REVIEWED IN THE PRODUCT EVALUATION AND NO NON-CONFORMANCES WERE FOUND FOR THIS LOT. THERE IS NO INDICATION OF MANUFACTURING DEFECTS. THIS IS REPORT 4 OF 6 FOR THE SAME EVENT. REPORTS 1, 2, 3, 5, AND 6 ARE REPORTED ON MFR # 0001032347-2016-00230-2, 0001032347-2016-00263-1, 0001032347-2016-00264-1, 0001032347-2016-00266-1, AND 0001032347-2016-00267-1.
IT WAS REPORTED THAT SIX ELEVATOR #301'S WERE BROKEN AND/OR BENT DURING DIFFERENT SURGERIES. THERE WAS NO DELAY IN ANY OF THE SURGERIES, ALL PARTS WERE RETRIEVED IN EACH CASE, AND THERE WAS NO INJURY TO THE PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350431 | ELEVATOR #301 | ELEVATOR | EMJ | BIOMET MICROFIXATION | N/A | 040207C07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |