FDA Adverse Event Malfunction Summary report: N

ELEVATOR #301

MDR report key: 5694752 · Received June 2, 2016

Report

Report Number
0001032347-2016-00267
Event Type
Malfunction
Date Received
June 2, 2016
Report Date
April 25, 2016
Manufacturer
BIOMET MICROFIXATION
Product Code
EMJ
PMA / PMN Number
PEXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE WARNINGS IN THE PACKAGE INSERT STATE THIS TYPE OF EVENT CAN OCCUR. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. A FOLLOW UP REPORT WILL BE SENT UPON COMPLETION OF THE DEVICE EVALUATION. THIS IS REPORT 5 OF 5 FOR THE SAME EVENT. REPORTS 1-4 ARE REPORTED ON MFR #:0001032347-2016-00263 THROUGH 0001032347-2016-00266.

Additional Manufacturer Narrative · 1

UDI #: (B)(4). REVIEW OF THE DEVICE HISTORY RECORDS SHOW THAT THE LOT WAS RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. ACCORDING TO THE PRODUCT EVALUATION, THE COMPLAINT IS CONFIRMED AS THE TIP OF THE ELEVATOR HAS FRACTURED. THE MOST-LIKELY, UNDERLYING CAUSE OF THE COMPLAINT WAS DETERMINED TO BE EXCESSIVE FORCE. THE INSTRUCTIONS FOR USE STATES, "THE TIP OF THE INSTRUMENT IS EXTREMELY THIN AND DELICATE; CARE SHOULD BE TAKEN TO AVOID APPLYING SIGNIFICANT PRESSURE TO THE TIP." THE NON-CONFORMANCE DATABASE WAS REVIEWED IN THE PRODUCT EVALUATION AND NO NON-CONFORMANCES WERE FOUND FOR THIS LOT. THERE IS NO INDICATION OF MANUFACTURING DEFECTS. THIS IS REPORT 6 OF 6 FOR THE SAME EVENT. REPORTS 1, AND 2 THROUGH 5 ARE REPORTED ON MFR # 0001032347-2016-00230-2, AND 0001032347-2016-00263-1 THROUGH 0001032347-2016-00266-1.

Description of Event or Problem · 1

IT WAS REPORTED THAT SIX ELEVATOR #301'S WERE BROKEN AND/OR BENT DURING DIFFERENT SURGERIES. THERE WAS NO DELAY IN ANY OF THE SURGERIES, ALL PARTS WERE RETRIEVED IN EACH CASE, AND THERE WAS NO INJURY TO THE PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349050 ELEVATOR #301 ELEVATOR EMJ BIOMET MICROFIXATION N/A 021216B16

Patients

Seq Age Sex Outcome Treatment
1