FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 5681323 · Received May 26, 2016

Report

Report Number
1314492-2016-03234
Event Type
Malfunction
Date Received
May 26, 2016
Date of Event
May 3, 2016
Report Date
May 4, 2016
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K133801
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF NO.: (B)(4) THE DEVICE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, AN EVALUATION COULD NOT BE COMPLETED. IF THE DEVICE IS RETURNED, AN EVALUATION WILL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND IN SPECIFICATION IN RELATION TO THE REPORTED FLOW RATE ERROR, WHICH WAS NOT REPRODUCED DURING FLOW RATE TESTING. BAXTER'S DEVICE EVALUATION FOUND THE DEVICE TO HAVE PASSING RESULTS WITH THE FOLLOWING TESTS: RATE = 40 ML/ HR, VTBI = 10 ML, DELIVERY = 9.578ML (-4.22%); RATE = 100 ML/HR, VTBI = 25 ML, DELIVERY = 24.046ML (-3.816%); RATE = 400 ML/HR, VTBI = 100 ML, DELIVERY = 94.771ML (-5.229%); RATE = 800 ML/HR, VTBI = 200 ML, DELIVERY = 192.112ML (-3.944%); RATE = 999 ML/HR, VTBI = 250 ML, DELIVERY = 241.133ML (-3.5468%). FLOWRATE DELIVERY WAS FOUND TO BE OUT OF THE ±5% ACCURACY AT VARIOUS RATES ON THE CUSTOMER SUPPLIED IV SET. NO COMPONENT COULD BE IDENTIFIED FOR CAUSALITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DELIVERED AT A LOWER RATE OR VOLUME THAN PROGRAMMED (UNDER INFUSION) DURING AN INFUSION OF VANCOMYCIN IN THE MEDICAL SURGICAL UNIT (PROGRAMMED AMOUNT AND DELIVERY RATE UNKNOWN). IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY. ALL THE INFORMATION REGARDING THE PATIENT AND EVENT REPORTED TO BAXTER BY THE CUSTOMER HAS BEEN REFLECTED IN THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337370 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1