FDA Adverse Event Injury Summary report: N

NONIN WRISTOX2 PULSE OXIMETER

MDR report key: 5676134 · Received May 24, 2016

Report

Report Number
3007389703-2016-00001
Event Type
Injury
Date Received
May 24, 2016
Date of Event
April 21, 2016
Report Date
April 21, 2016
Manufacturer
NONIN MEDICAL INC.
Product Code
DQA
PMA / PMN Number
K102350
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PATIENT DATA TRACEABILITY RELYING ON (B)(4) FILES AND (B)(4) IDENTIFICATION USED. (B)(4) IS A REPACKAGER OF THE NONIN 3150 PULSE OXIMETER, THAT IS INCLUDED IN THE NOX T3 SYSTEM. THE LEGAL MANUFACTURER OF THE FINISHED DEVICE IS NONIN MEDICAL INC. THAT HAS BEEN NOTIFIED OF THE REPORTED EVENT. THE SUSPECTED DEVICE HAS BEEN SENT DIRECTLY FROM (B)(4) (THE IMPORTER AND DISTRIBUTOR) TO NONIN MEDICAL (THE FINISHED DEVICE LEGAL MANUFACTURER) FOR FURTHER EVALUATION. (B)(4) HAS EVALUATED THE INFORMATION RELATED TO THIS REPORTED EVENT AND CAME TO THE CONCLUSION THAT THE INCIDENT IS NOT RELATED TO THE REPACKAGING OF THE NONIN 3150 PULSE OXIMETER PERFORMED BY (B)(4). ANY FURTHER REPORTS WILL COME FROM NONIN MEDICAL THE FINISHED DEVICE MANUFACTURER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ONE OF THEIR PATIENTS COMPLAINED OF A BURNING SENSATION AFTER WEARING THE NONIN 3150 PULSE OXIMETER FOR APPROXIMATELY TWO HOURS DURING AN IN HOME SLEEP STUDY. THE PATIENT REMOVED THE DEVICE AND FOUND THAT THE AREA ON THE PATIENT'S SKIN UNDER WERE THE BATTERIES ARE INSERTED HAD REDNESS AND A BLISTER. THE PATIENT WENT TO URGENT CARE AND WAS TREATED WITH AN ANTIBIOTIC AND A CREAM FOR THE BURN. THE DEVICE WAS RETURNED TO THE CUSTOMER BY THE PATIENT AND THE PATIENT WAS SENT TO A CLINIC FOR THE COMPLETION OF THE SLEEP STUDY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332292 NONIN WRISTOX2 PULSE OXIMETER WRIST-WORN PULSE OXIMETER DQA NONIN MEDICAL INC. MODEL 3150

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other NOX T3 SYSTEM SN: (B)(4)