I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2016-00029
- Event Type
- Malfunction
- Date Received
- May 20, 2016
- Date of Event
- April 27, 2016
- Report Date
- July 7, 2016
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4)
APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 06/22/2016. CUSTOMER RETURNS AND RETAIN PRODUCT WAS TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
ON (B)(6) 2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT RESULTS ON A (B)(6) MALE PATIENT ADMITTED FOR VOMITING. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. THE CUSTOMER STATES THAT THE PATIENT WAS TREATED ON THE REPEAT OF THE I-STAT RESULTS AND PRODUCT IS AVAILABLE FOR INVESTIGATION. DATE SAMPLE TIME TEST TIME NA CL SAMPLE DATE: (B)(6)2016, SAMPLE TIME: 0240, TEST TIME: 0246, NA: 120, CL: >140, SAMPLE: A.; ON (B)(6) 2016, N/A, N/A, 143, 104, N/A. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.
NA
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 325353 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H16008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |