FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 5668204 · Received May 19, 2016

Report

Report Number
1052693-2016-00966
Event Type
Malfunction
Date Received
May 19, 2016
Date of Event
April 20, 2016
Report Date
May 19, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. RETURNED TEST STRIPS EVALUATED WITH DEFECT FOUND. MOST LIKELY ROOT CAUSE OF PQQ-PD LOT PS2582 WITH LOW GLUCOSE READINGS IS DUE TO RETURNED STRIPS HAD BEEN LEFT OUTSIDE OF THE VIAL FOR A LONG PERIOD TIME WHICH ABSORBED MOISTURE AS A RESULT OF LOW METER GLUCOSE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR LOW BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 113, 101MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120-150MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 11/21/2018 AND OPEN VIAL DATE IS 4/13/2016 .THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6). MEMORY CONCERNS: UNDISCLOSED. CALLED CUSTOMER ABOUT THE METER, SHE SAID THAT THE METER IS READING LOW, COMPARED TO HER NORMAL; SHE NORMALLY DOES NOT RUN THIS LOW. THE RESULTS ABOVE WERE A FASTING 113MG/DL AND THEN 2 HOURS AFTER EATING A HAMBURGER AND RESULTS WERE 101MG/DL. HER NORMAL WAS BASED ON A DIFFERENT METER SHE HAD BEEN USING BUT ABOUT 1 YEAR AGO THE DR'S BLOOD RESULTS WERE AROUND 140MG/DL FOR A FASTING. NOTHING HAS CHANGED AS FAR AS MEDICATION EXCEPT WITH THIS NEW METER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321431 TRUERESULT BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUERESULT PS2582

Patients

Seq Age Sex Outcome Treatment
1 0 YR