FDA Adverse Event
Malfunction
Summary report: N
GRAB N' GO VANTAGE VALVE INTEGRATED PRESSURE REGULATOR FOR D OR E CYLINDER COMPA
MDR report key: 5662241
·
Received May 13, 2016
Report
- Report Number
- MW5062337
- Event Type
- Malfunction
- Date Received
- May 13, 2016
- Date of Event
- May 4, 2016
- Report Date
- May 13, 2016
- Manufacturer
- PRAXAIR DISTRIBUTION, INC.
- Product Code
- CAN
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
A GRAB N' GO VANTAGE VALVE INTEGRATED PRESSURE REGULATOR WAS IN PLACE ON AN ALUMINUM OXYGEN CYLINDER MARKED MR COMPATIBLE. THE TECH NOTED A PULL OF THE CYLINDER FROM REGULATOR TOWARDS THE 1.5 TESLA MRI MAGNET AND WAS ABLE TO PREVENT ITEM FROM BEING DRAWN INTO MAGNET. PRAXAIR DISTRIBUTION, INC., LOGGED THE (B)(4). THE COMPANY USES SHELLOCK MRI SERVICES, INC., TO EVALUATE THEIR PRODUCTS BASED ON ASTMF2503-8 SPECIFICATIONS. THE PRAXAIR AIR GRAB N' GO UNIT WAS TESTED FOR TRANSLATIONS ATTRACTION AT 3 TESLA. KNOWN FERROUS METAL PIN OR C-CLIP. STATES SHOULD NOT HAVE EXPERIENCED PULL. THE TECH NOTICED THE PULL AS ENTERED ZONE 4 OF THE MAGNET ROOM OF 1.5 TESLA. NO INJURY OR DAMAGE TO STAFF, PATIENTS OR MAGNET OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307518 | GRAB N' GO VANTAGE VALVE INTEGRATED PRESSURE REGULATOR FOR D OR E CYLINDER COMPA | OXYGEN REGULATOR | CAN | PRAXAIR DISTRIBUTION, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |