AEROAUTOCAT 2 WAVE
Report
- Report Number
- 1219856-2016-00117
- Event Type
- Malfunction
- Date Received
- May 17, 2016
- Date of Event
- April 27, 2016
- Report Date
- April 27, 2016
- Manufacturer
- ARROW INTERNATIONAL INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
QN#(B)(4). THE PCS ASSEMBLY (P/N 96-3006-001, S/N (B)(4)) WAS RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE PCS ASSEMBLY WAS PERFORMED AND NO ABNORMALITIES WERE NOTED. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED INTO A KNOWN GOOD AUTOCAT2W AND PERFORMED FUNCTIONAL TESTING. THE PURGE CYCLE WAS PERFORMED MULTIPLE TIMES AND PASSED EACH TIME WITHOUT ALARM; HOWEVER, THE PCS ASSEMBLY FAILED HELIUM LEAK TEST (PATIENT SIDE), THE BALLOON BASELINE PRESSURE DROPPED 6 MMHG APPROXIMATELY 15 MINUTES AFTER THE PUMP WAS INITIATED PUMPING. THE KNOWN GOOD AUTOCAT2W (ENG081202W)WITH THE PCS ASSEMBLY IN QUESTION INSTALLED FAILED THE FUNCTIONAL TEST. THE PCS ASSEMBLY WAS THEN REMOVED FROM THE PUMP. THE PCS ASSEMBLY IN QUESTION WAS INSTALLED ONTO THE LEAK TESTER AND FAILED LEAK TESTING. A SMALL LEAK WAS NOTED TO BE COMING FROM THE DRAIN PORT. VISUAL INSPECTION OF THE PCS ASSEMBLY INTERNAL HARDWARE WAS PERFORMED. RUST AND DRIED CONDENSATIONS WERE FOUND INSIDE THE DRAIN VALVE AND MANIFOLD BLOCK WHERE THE DRAIN VALVE WAS CONNECTED. NOTICE: THE PURGE FAILURE ALARM WOULD OCCUR IF SYSTEM COULD NOT PULL A NEGATIVE PRESSURE (DRAIN VALVE). OTHER REMARKS: A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE LOT NUMBER/SERIAL NUMBER WITH NO RELEVANT FINDINGS. THE DEVICE PASSED ALL MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE.
(B)(4)
IT WAS REPORTED PER FIELD SERVICE REPORT: SYMPTOM - PURGE FAILURE (5). FINDINGS/ACTION TAKEN: CONFIRMED FAILURE. COULD NOT PULL NEGATIVE (DRAIN VALVE). REPLACED PCS (PNEUMATIC CONTROL SWITCH); ALL WORKS, BUT HAS A SMALL HELIUM LOSS TRACED TO INTERFACE BLOCK. REPLACED INTERFACE BLOCK. FUNCTIONAL CHECK WAS PERFORMED; PASS ALL. SOFTWARE LEVEL: 2.24. ADDITIONAL INFORMATION STATED THAT THE TRANSPORT TEAM ARRIVED AND THE PATIENT WAS ON A COMPETITOR'S INTRA-AORTIC BALLOON PUMP (IABP). WHEN THE TEAM SWAPPED OUT THE PUMP THE ARROW PUMP ALARMED PURGE FAILURE (5). AS A RESULT, THE PATIENT WAS PUT BACK ON THE COMPETITOR'S PUMP AND A SECOND TRANSPORT TEAM WAS CALLED. THE SECOND TEAMS ARROW PUMP WORKED FINE AND THE PATIENT WAS TRANSPORTED TO THE NEXT FACILITY. THERE WAS A REPORTED DELAY IN TRANSPORT, NOT IN IABP THERAPY. IT IS UNKNOWN IF THE DELAY / INTERRUPTION WAS DETRIMENTAL TO THE PATIENT'S HEALTH PLAN.
IT WAS REPORTED PER FIELD SERVICE REPORT: SYMPTOM - PURGE FAILURE (5). FINDINGS/ACTION TAKEN: CONFIRMED FAILURE. COULD NOT PULL NEGATIVE (DRAIN VALVE). REPLACED PCS (PNEUMATIC CONTROL SWITCH); ALL WORKS , BUT HAS A SMALL HELIUM LOSS TRACED TO INTERFACE BLOCK. REPLACED INTERFACE BLOCK. FUNCTIONAL CHECK WAS PERFORMED; PASS ALL. SOFTWARE LEVEL: 2.24. ADDITIONAL INFORMATION STATED THAT THE TRANSPORT TEAM ARRIVED AND THE PATIENT WAS ON A COMPETITOR'S INTRA-AORTIC BALLOON PUMP (IABP). WHEN THE TEAM SWAPPED OUT THE PUMP THE ARROW PUMP ALARMED PURGE FAILURE (5). AS A RESULT, THE PATIENT WAS PUT BACK ON THE COMPETITOR'S PUMP AND A SECOND TRANSPORT TEAM WAS CALLED. THE SECOND TEAMS ARROW PUMP WORKED FINE AND THE PATIENT WAS TRANSPORTED TO THE NEXT FACILITY. THERE WAS A REPORTED DELAY IN TRANSPORT, NOT IN IABP THERAPY. IT IS UNKNOWN IF THE DELAY / INTERRUPTION WAS DETRIMENTAL TO THE PATIENT'S HEALTH PLAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316182 | AEROAUTOCAT 2 WAVE | INTRA-AORTIC BALLOON PUMP | DSP | ARROW INTERNATIONAL INC. | 40540W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |