FDA Adverse Event Malfunction Summary report: N

OT VERIO2 METER

MDR report key: 5658787 · Received May 17, 2016

Report

Report Number
2939301-2016-24131
Event Type
Malfunction
Date Received
May 17, 2016
Report Date
April 27, 2016
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K131363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING THAT THE SUBJECT METER READ INACCURATELY ERRATIC. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF "156, 101, 134 AND 121 MG/DL" WITH THE SUBJECT METER, PERFORMED WITHIN 20 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR PRECISION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316737 OT VERIO2 METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3986652

Patients

Seq Age Sex Outcome Treatment
1