FDA Adverse Event
Injury
Summary report: N
CENTURION VISION SYSTEM
MDR report key: 5656054
·
Received May 16, 2016
Report
- Report Number
- 2028159-2016-02025
- Event Type
- Injury
- Date Received
- May 16, 2016
- Date of Event
- April 13, 2016
- Report Date
- May 14, 2016
- Manufacturer
- ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K121555
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
A DOCTOR REPORTED A PATIENT EXPERIENCED POST OPERATIVE INFLAMMATION RESEMBLING TOXIC ANTERIOR SEGMENT SYNDROME (TASS) FOLLOWING A CATARACT EXTRACTION PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THE SITE CHANGED THEIR INTERNAL PROTOCOL FOR CLEANING THE IRRIGATION/ASPIRATION HANDPIECES AND HAVE HAD NO CASES OF TASS SINCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 311294 | CENTURION VISION SYSTEM | UNIT, PHACOFRAGMENTATION | HQC | ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER | NA | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DUOVISC |