ANEURX STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2016-01020
- Event Type
- Injury
- Date Received
- May 13, 2016
- Date of Event
- May 16, 2002
- Report Date
- April 28, 2016
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
DATE OF EVENT: THE EXACT DATES OF THE EVENTS ARE UNKNOWN. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
CORRECTED INFORMATION: SEX. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED THE FOLLOWING INFORMATION OBTAINED FROM THE JOURNAL ARTICLE ENTITLED; ANEURYSM NECK DIAMETER AFTER ENDOVASCULAR REPAIR OF ABDOMINAL AORTIC ANEURYSMS MOHAMMAD F. BADRAN, MD, DEREK A. GOULD, MD, IRFAN RAZA, RICHARD G. MCWILLIAMS, MD, OLIVER BROWN, BA, PETER L. HARRIS, MD, GEOFFREY L. GILLING-SMITH, MD, JOHN BRENNAN, MD, DONNA WHITE, DCR(R), SHAUN MEAKIN, SRN, AND PETER C. ROWLANDS, MD (J VASC INTERV RADIOL 2002; 13:887¿892) ANEURX AND TALENT STENT GRAFT SYSTEMS WERE IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF ABDOMINAL ANEURYSM REPAIR. THE FOLLOWING ADVERSE EVENTS WERE OBSERVED: TYPE IA ENDOLEAK, OPEN REPAIR. PURPOSE: TO DETERMINE ANEURYSM NECK DIAMETER CHANGE AFTER ENDOVASCULAR REPAIR (EVR) OF ABDOMINAL AORTIC ANEURYSM (AAA) AND ITS RELATIONSHIP TO STENT-GRAFT DIAMETER. MATERIALS AND METHODS: THE CASES OF 73 PATIENTS WITH AAAS WHO UNDERWENT EVR WERE REVIEWED RETROSPECTIVELY: 68 HAD PREOPERATIVE IMAGING AVAILABLE. NECK DIAMETER WAS REVIEWED BY A SINGLE OBSERVER (M.F.B.) ON PREOPERATIVE, IMMEDIATE POSTOPERATIVE, ANNUAL, AND MOST RECENT CONTRAST-ENHANCED COMPUTED TOMOGRAPHIC SCANS. BASELINE AND FOLLOW-UP NECK DIAMETERS WERE COMPARED WITH THE MANUFACTURERS¿ VALUES FOR UNCONSTRAINED STENT-GRAFT DIAMETERS. RESULTS: INTRAOBSERVER ERROR WAS 2 MM. ANEURYSM NECK DIAMETER INCREASED FROM 21.8 MM (RANGE, 17¿28 MM) AT BASELINE TO 22.8 MM (RANGE, 19¿30 MM) POSTOPERATIVELY AND 25.8 MM (RANGE, 19¿31 MM; P <(><<)> .001) AT A MEAN FOLLOW-UP OF 25.5 MONTHS (RANGE, 6.2¿ 60.8 MO). NECK DIAMETER INCREASE WAS MORE THAN 2 MM IN 24 PATIENTS (33%). MEAN CHANGE IN THE FIRST, SECOND, THIRD, AND FOURTH YEARS WAS 1.63 MM, 0.52 MM, 0.25 MM, AND 0.33 MM, RESPECTIVELY. BASELINE MEAN STENT-GRAFT OVERSIZING WAS 2.9 MM (13.7%; RANGE, - 1 TO 8 MM), WHICH DECREASED TO 0.7 MM (RANGE, - 4 TO 6 MM) AT LATEST FOLLOW-UP. NECK DIAMETER EXCEEDED STENT-GRAFT DIAMETER (MEAN, 1.8 MM; RANGE, 1¿4 MM) IN 21 CASES (28%) AND BY MORE THAN 2 MMIN FIVE CASES (6.8%). WHEN NECK DIAMETER CHANGE WAS CORRELATED WITH CHANGE IN SAC DIAMETER, IT WAS FOUND TO BE INSIGNIFICANT (P = .24); HOWEVER, IT WAS SIGNIFICANTLY CORRELATED WITH BASELINE OVERSIZING (P = .01). CONCLUSIONS: AFTER EVR, THE ANEURYSM NECK DILATES, MOSTLY IN THE FIRST 2 YEARS, BY GREATER THAN 2 MM IN ONE THIRD OF PATIENTS. THIS IS POSSIBLY RELATED TO THE PRESENCE OF THE ENDOGRAFT. THE ASSOCIATED REDUCTION OF STENT-GRAFT OVERSIZING WARRANTS CONTINUED VIGILANCE FOR PROXIMAL ENDOLEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309512 | ANEURX STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention |