FDA Adverse Event Malfunction Summary report: N

CAREBRUSH

MDR report key: 5652832 · Received May 13, 2016

Report

Report Number
1122376-2016-00012
Event Type
Malfunction
Date Received
May 13, 2016
Date of Event
March 29, 2016
Report Date
May 12, 2016
Manufacturer
QIAGEN
Product Code
HHT
PMA / PMN Number
K971586
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS BEING USED IN (B)(6) AS PART OF A CUSTOMER'S STUDY FOR SELF SAMPLE COLLECTION. THE CURRENTLY MARKETED INTENDED USE FOR CAREBRUSH IS TO BE USED BY PHYSICIANS AND TRAINED MEDICAL PROFESSIONALS FOR THE COLLECTION OF CERVICAL SPECIMENS FOR SUBSEQUENT TESTING OF THE (B)(6) TEST. SEVERAL ATTEMPTS WERE MADE TO FOLLOW UP WITH THE REPORTING FACILITY TO OBTAIN FURTHER DETAILS AROUND THIS INCIDENT. THE REPORTING FACILITY WAS UNABLE TO PROVIDE FURTHER DETAILS. SIMILAR COMPLAINTS HAVE BEEN REPORTED FOR THIS BRUSH. PRIOR INVESTIGATION CONCLUDED THAT THE DEVICE MALFUNCTION WAS DUE TO THE BRUSH MISSING THE HEATING PROCESS WHICH ADHERES IT TO THE STEM. THERE HAVE BEEN A TOTAL OF 5 COMPLAINTS FOR DETACHED BRUSH TIPS REPORTED SINCE THIS DEVICE HAS BEEN IN DISTRIBUTION (1997). OVER (B)(4) BRUSHES HAVE BEEN DISTRIBUTED. (B)(4). THIS DEFECT IS NOT LIKELY TO CAUSE AN ADVERSE EVENT AND THE PROBABILITY OF THIS DEFECT REQUIRING MEDICAL OR SURGICAL INTERVENTION IS REMOTE. THIS PRODUCT IS NOT APPROVED OR DISTRIBUTED IN THE US, HOWEVER THE BRUSH IS SUBSTANTIALLY SIMILAR TO THE DIGENE HC2 DNA COLLECTION DEVICE THAT IS APPROVED IN THE US. DEVICE WAS NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

ON 13APR2016 QIAGEN RECEIVED A REPORT, (B)(4), FROM (B)(6) OF AN INCIDENT WHERE THE BRUSH HEAD BECAME DETACHED FROM THE SHAFT AND REMAINED INSIDE THE END USER. THE NURSING ASSISTANT EXAMINED THE END USER AND COULD NOT LOCATE THE BRUSH. TWO DAYS LATER THE END USER RETURNED TO THE HEALTH FACILITY AND INFORMED THEM THAT THE BRUSH WAS EXPULSED AND NO INJURIES WERE INCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308800 CAREBRUSH CAREBRUSH HHT QIAGEN UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 25 YR