COBE CENTRYSTEM 3
Report
- Report Number
- 1713683-1996-00263
- Event Type
- Malfunction
- Date Received
- December 17, 1996
- Report Date
- December 16, 1996
- Manufacturer
- GAMBRO HEALTHCARE
- Product Code
- FII
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
RETUNRED FOUR-POSITION LINE CLAMP WAS VISUALLY INSPECTED AND CONFIRMED THAT CLAMP HOLDER WAS BROKEN. THE TABS ON THE WHITE CLAMP INSERT HOLDER WAS BROKEN OR STRETCHED, ALLOWING THE INSERTS TO FALL OUT OF THE HOLDER. IT WAS ALSO DETERMINED THAT THE RETUNRED CLAMP STOP TABS WHICH HOLD THE CLAMP IN THE OCCLUDED POSITION WERE EITHER WORN OR BROKEN, SO THAT THE CLAMP NO LONGER STOPPED SECURELY IN THE CLOSED POSITION. THIS CONDITION OCCURRED WITH NORMAL USE. THE ABOVE INFO INDICATED THAT THE LINE CLAMP FAILURE DESCRIBED IN THIS COMPLAINT WAS GENERATED BY A BROKEN CLAMP ASSEMBLY. THIS ISSUE WILL CONTINUE TO BE TRENDED AS PART OF THE GAMBRO HEALTHCARE CORRECTIVE ACTION SYSTEM AND THE MANAGEMENT REVIEW TEAM. ANY FURTHER INVESTIGATIONS, ANALYSES, AND/OR CORRECTIVE ACTION TAKEN ON THIS COMPLAINT ISSUE WILL BE DETERMINED BY AND DOCUMENTED IN THIS CORRECTIVE ACTION REVIEW PROCESS. SEE FAR#960027. CUSTOMER CONTACTED: NO. SALES/SERVICE CONTACTED BY INVESTIGATOR? NO. TRAINING REQUIRED? NO.
THE LIP THAT HOLDS THE CAM IS EITHER WORN AND/OR BROKE OFF AND THE CAMS ARE FALLING OUT OR NOT OCCLUDING THE TUBING SET. THERE WERE NO PT INJURIES OR MEDICAL INTERVENTIONS REPORTED. THERE WERE NO RELATED CPF DEFECTS REPORTED FOR THIS MACHINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBE CENTRYSTEM 3 | DIALYSIS CONTROL UNIT | FII | GAMBRO HEALTHCARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |