INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2016-00208
- Event Type
- Malfunction
- Date Received
- May 13, 2016
- Date of Event
- April 11, 2016
- Report Date
- April 13, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K103767
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT265 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4). THE RETURNED BREATHING CIRCUIT WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE PROXIMAL CONNECTOR OF THE EXPIRATORY LIMB CRACKED. NO OTHER DAMAGE WAS OBSERVED TO THE BREATHING CIRCUIT. THE PRESSURE TEST SHOWED THAT THE RETURNED BREATHING CIRCUIT WAS WITHIN SPECIFICATION. A LOT CHECK REVEALED THREE OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 150914. CONCLUSION: WE ARE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE DAMAGE NOTED ON THE RETURNED BREATHING CIRCUIT. PREVIOUS INVESTIGATIONS OF SIMILAR COMPLAINTS SUGGEST THAT THE OBSERVED CRACKING ON THE CONNECTOR WAS MOST LIKELY DUE TO THE CONNECTOR COMING INTO CONTACT WITH CLEANING CHEMICALS RESULTING IN ENVIRONMENTAL STRESS CRACKING. ALL RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THE HOSPITAL REPORTED THAT THE DAMAGE OCCURRED AFTER THREE DAYS OF USE, WHICH SUGGESTS THAT THE COMPLAINT CIRCUIT BECAME DAMAGED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT265 STATE THE FOLLOWING: CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. THE USER INSTRUCTIONS ALSO STATE THE FOLLOWING: DO NOT SOAK, WASH, STERILIZE OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS OR HAND SANITIZERS.
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT THEY FOUND A CRACK ON THE COLLAR ON THE EXPIRATORY LIMB OF AN RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT. THIS WAS FOUND AFTER THREE DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309492 | INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT265 | 150914 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |