FDA Adverse Event Malfunction Summary report: N

COBE CENTRYSYSTEM 3

MDR report key: 56523 · Received December 17, 1996

Report

Report Number
1713683-1996-00259
Event Type
Malfunction
Date Received
December 17, 1996
Report Date
December 17, 1996
Manufacturer
GAMBRO HEALTHCARE
Product Code
FII
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE WERE NO RELATED CPF DEFECTS REPORTED FOR THIS MACHINE. RETUREND FOUR-POSITION LINE CLAMP WAS VISUALLY INSPECTED AND CONFIRMED THAT CLAMP HOLDER WAS BROKEN. TABS ON WHITE CLAMP INSERT HOLDER WAS BROKEN OR STRETCHED, ALLOWING INSERTS TO FALL OUT OF HOLDER. IT WAS ALSO DETERMINED THAT RETURNED CLAMP STOP TABS WHICH HOLD CLAMP IN OCCLUDED POSITION WERE EITHER WORN OR BROKEN, SO THAT CLAMP NO LONER STOPPED SECURELY IN CLOSED POSTION. THIS CONDITION OCCURRED WITH NORMAL USE. ABOVE INFO INDICATES THAT LINE CLAMP FAILURE DESCRIBED IN THIS COMPLAINT WAS GENERATED BY A BROKEN CLAMP ASSEMBLY. THIS ISSUE WILL CONTINE TO BE TRENDED AS PART OF GAMBRO HEALTHCARE CORRECTIVE ACTION SYSTEM AND MANAGEMENT REVIEW TEAM. ANY FURTHER INVESTIGATIONS, ANALYSES, AND/OR CORRECTIVE ACTIONS TAKEN ON THIS COMPLAINT ISSUE WILL BE DETERMINED BY AND DOCUMENTED IN THIS CORRECTIVE ACTION REVIEW PROCESS. SEE FAR# 960027. CUSTOMER CONTACTED: NO. SALES/SERVICE CONTACTED BY INVESTIGATOR? NO. TRAINING REQUIRED? NO.

Description of Event or Problem · 1

DURING A DIALYSIS TREATMENT, A FOUR POSITION LINE CLAMP BROKE. THERE WAS NO PT INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBE CENTRYSYSTEM 3 DIALYSIS CONTROL UNIT FII GAMBRO HEALTHCARE NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other