ANEURX STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2016-00986
- Event Type
- Injury
- Date Received
- May 11, 2016
- Date of Event
- February 23, 2003
- Report Date
- April 26, 2016
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA ROSA
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE EXACT DATES OF THE EVENTS ARE UNKNOWN.
CORRECTED INFORMATION: SEX. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED THE FOLLOWING INFORMATION OBTAINED FROM THE JOURNAL ARTICLE ENTITLED; MORPHOLOGIC CHANGES AND OUTCOME FOLLOWING ENDOVASCULAR ABDOMINAL AORTIC ANEURYSM REPAIR AS A FUNCTION OF ANEURYSM SIZE. FRANK R. ARKO, MD; KONSTANTINO A. FILIS, MD; BRADLEY B. HILL, MD; THOMAS J. FOGARTY, MD; CHRISTOPHER K. ZARINS, MD (ARCH SURG. 2003; 138:651-656). AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF ABDOMINAL ANEURYSM REPAIR. IT WAS REPORTED THAT A LATE SECONDARY INTERVENTION WAS PERFORMED DUE TO THE PROXIMAL AORTIC NECK MEASUREMENT OF 30 MM IN DIAMETER ON THE CTA AND 26 MM ON THE INTRAVASCULAR ULTRASOUND WITH AN OVERALL NECK LENGTH OF 10 MM. IT WAS REPORTED THAT THE PATIENT HAD PERSISTENT TYPE I ENDOLEAK FOLLOWING INDEX PROCEDURE, WITHOUT AN INCREASE IN ANEURYSM DIAMETER AT 18 MONTHS. SURGICAL CONVERSION WAS DONE. THE PATIENT WAS SUCCESSFULLY TREATED WITHOUT COMPLICATIONS. IT WAS NOTED THAT THE SURGICAL CONVERSION WAS A RESULT OF POOR PATIENT SELECTION; THE PATIENT HAD POOR PROXIMAL AORTIC NECK ANATOMY PREOPERATIVELY, WITH SHORT AND ANGULATED NECK. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE. HYPOTHESIS SMALL INFRARENAL ABDOMINAL AORTIC ANEURYSMS HAVE A MORE FAVORABLE CLINICAL AND MORPHOLOGIC OUTCOME COMPARED WITH MEDIUM AND LARGE ABDOMINAL AORTIC ANEURYSMS FOLLOWING ENDOVASCULAR ANEURYSM REPAIR(EVAR). DESIGN A PROSPECTIVE CLINICAL SERIES OF 206 PATIENTS UNDERGOING ELECTIVE EVAR BETWEEN 1996 AND 2001. SETTING A TERTIARY CARE ACADEMIC HEALTH CENTER. PATIENTS WERE GROUPED ACCORDING TO ANEURYSM SIZE: SMALL (<(> <<)>50 MM), MEDIUM (50-60 MM), AND LARGE (>60 MM). INTERVENTIONS PRIMARY EVAR AND SECONDARY PROCEDURES TO SECURE FIXATION OF THE STENT GRAFT AND SURGICAL CONVERSIONS. MAIN OUTCOME MEASURES ANEURYSM DIAMETER, ENDOLEAKS, AND LONG-TERM MORPHOLOGIC CHANGES WERE ANALYZED POSTOPERATIVELY WITH 3-DIMENSIONAL RECONSTRUCTIONS OF COMPUTED TOMOGRAPHIC ANGIOGRAMS. RESULTS GROUPS WERE SIMILAR IN AGE, COMORBIDITIES, AND FOLLOW-UP (MEAN ± SD, 32.1 ± 11.8 MONTHS). THERE WERE 30 SMALL ANEURYSMS, 92 MEDIUM ANEURYSMS, AND 84 LARGE ANEURYSMS, WITH A MEAN SIZE OF 45.1 ± 3.7 MM, 53.8 ± 3.1 MM, AND 66.1 ± 6.8 MM, RESPECTIVELY (P<(><<)>.01). THERE WAS NO SIGNIFICANT DIFFERENCE IN PROXIMAL NECK OR ILIAC ARTERY DIAMETER AMONG THE 3 GROUPS. THE PROXIMAL AORTIC NECK LENGTH (28.1 ± 11.6 MM [SMALL]; 23.9 ± 11.3 MM [MEDIUM]; AND 22.1 ± 11.6 MM [LARGE]; P<(> <<)>.05) WAS SIGNIFICANTLY SHORTER IN LARGE ANEURYSMS. FURTHERMORE, THERE WAS A SIGNIFICANT INCREASE (6% [SMALL]; 15% [MEDIUM]; AND 21% [LARGE]; P<(><<)>.05) IN ANGULATED NECKS IN LARGE ANEURYSMS. FOLLOWING TREATMENT, ANEURYSM DIAMETER REMAINED STABLE IN MOST PATIENTS (83% [SMALL]; 82% [MEDIUM]; AND 83% [LARGE]), WITH A MEAN DECREASE OF 2.0 ± 6.5 MM, 2.1 ± 6.1 MM, AND 3.7 ± 7.7 MM IN EACH GROUP, RESPECTIVELY (P = .45). THERE WAS NO DIFFERENCE IN THE INCIDENCE OF ENDOLEAKS, ANEURYSM CONTRACTION, OR ANEURYSM EXPANSION BASED ON PREOPERATIVE ANEURYSM DIAMETER. SECONDARY PROCEDURES WERE PERFORMED IN 5 (20%) OF 25, 9 (5.2%) OF 170, AND 5 (36%) OF 11 ANEURYSMS THAT CONTRACTED, REMAINED STABLE, OR EXPANDED, RESPECTIVELY, FOLLOWING EVAR (P<(><<)>.05). CONCLUSIONS THERE IS A 15% INCREASE IN NECK ANGULATION AND A 27% DECREASE IN NECK LENGTH IN LARGE COMPARED WITH SMALL INFRARENAL ABDOMINAL AORTIC ANEURYSMS, WITH NO DIFFERENCE IN OUTCOME. ANEURYSMS THAT ARE STABLE FOLLOWING EVAR HAVE A SIGNIFICANTLY LOWER INCIDENCE OF REQUIRING SECONDARY PROCEDURES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300951 | ANEURX STENT GRAFT SYSTEM | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC CARDIOVASCULAR SANTA ROSA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |