GOLVO 7007 ES SANS ACCOUDOIRS
Report
- Report Number
- 8030916-2016-00063
- Event Type
- Malfunction
- Date Received
- May 11, 2016
- Date of Event
- April 26, 2016
- Report Date
- April 26, 2016
- Manufacturer
- LIKO AB
- Product Code
- FSA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
THE ACCOUNT FOUND THE GEAR RACK OF THE LIFT WAS BROKEN. IN THE SERVICE MANUAL FOR GOLVO 3EN400404, UNDER THE SECTION INSTRUCTIONS REGARDING THE CHECK POINTS, 5B BASE OPENING & CLOSING LINKAGE IT IS STATED: INSPECT GEAR RACK FOR WORN TEETH AND CRACKS WHERE STOP NUT CONTACTS BACK SURFACE. INSPECT FRICTION PLATES. REPLACE IF WORN OR DAMAGED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS LIFT. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR LIFTS. THE ACCOUNT REPLACED THE GEAR RACK SET TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING ONE LEG CAME OUT ALL THE WAY FROM THEIR PATIENT LIFT. THE LIFT WAS LOCATED IN THE PATIENT AREA AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).