FDA Adverse Event
Other
Summary report: N
BD LOGIC BLOOD GLUCOSE MONITOR
MDR report key: 564730
·
Received January 6, 2005
Report
- Report Number
- 2243072-2005-00008
- Event Type
- Other
- Date Received
- January 6, 2005
- Date of Event
- December 16, 2004
- Report Date
- January 6, 2005
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO NUMEROUS ERRONEOUS READINGS WITH THEIR METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (282) VS. BD READINGS OF 521, 408, 166, 161, 155 YIELDED DATA POINTS IN THE C/D ZONE OF THR GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |