VERSAFITCUP CC FLAT PE HC LINER Ø 36/G
Report
- Report Number
- 3005180920-2016-00213
- Event Type
- Injury
- Date Received
- May 11, 2016
- Date of Event
- April 10, 2016
- Report Date
- July 26, 2016
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- PMA / PMN Number
- K103352
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
ON 08 JULY 2016 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 18 JULY 2016 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
ON 29 APRIL 2016 THE MEDICAL AFFAIRS PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: RECURRING EARLY DISLOCATION IN A CEMENTLESS PRIMARY THA PERFORMED TWO MONTHS BEFORE. LATERALIZATION AND LENGTHENING HAS BEEN ACHIEVED THROUGH A MODIFIED HEAD AND SUBSTITUTION OF INLAY WITH A HOODED VERSION. THE COMPONENTS LOOK APPROPRIATELY POSITIONED, PROBABLY THE CAUSE FOR THE RECURRING DISLOCATION WAS INSUFFICIENT SOFT TISSUE TENSION OR SLIGHT IMMEDIATE SUBSIDENCE. NO REASON TO SUSPECT THAT A MALFUNCTIONING DEVICE CAUSED THE PROBLEM. BATCH REVIEW PERFORMED ON 09 MAY 2016. LOT 134964: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12 DECEMBER 2013. EXPIRATION DATE: 2018-10-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE L + 4, CODE 01.29.210, LOT. 152618 (K112115) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26 OCTOBER 2015. EXPIRATION DATE: 2020-10-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VERSAFITCUP CC TRIO ACETABULAR SHELL Ø 62, CODE 01.26.45.0062, LOT. 144699 (K103352) [NOT EXPLANTED] (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 OCTOBER 2014. EXPIRATION DATE: 2019-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT AVAILABLE.
THE PATIENT HAD 2 DISLOCATIONS AFTER THE PRIMARY SURGERY, SO THE SURGEON DECIDED TO REVISE HEAD AND LINER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 302891 | VERSAFITCUP CC FLAT PE HC LINER Ø 36/G | FLAT PE LINER | LZO | MEDACTA INTERNATIONAL SA | 134964 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |