FDA Adverse Event Injury Summary report: N

VERSAFITCUP CC FLAT PE HC LINER Ø 36/G

MDR report key: 5647013 · Received May 11, 2016

Report

Report Number
3005180920-2016-00213
Event Type
Injury
Date Received
May 11, 2016
Date of Event
April 10, 2016
Report Date
July 26, 2016
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K103352
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON 08 JULY 2016 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON 18 JULY 2016 THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.

Additional Manufacturer Narrative · 1

ON 29 APRIL 2016 THE MEDICAL AFFAIRS PERFORMED A CLINICAL EVALUATION AND COMMENTED AS FOLLOWS: RECURRING EARLY DISLOCATION IN A CEMENTLESS PRIMARY THA PERFORMED TWO MONTHS BEFORE. LATERALIZATION AND LENGTHENING HAS BEEN ACHIEVED THROUGH A MODIFIED HEAD AND SUBSTITUTION OF INLAY WITH A HOODED VERSION. THE COMPONENTS LOOK APPROPRIATELY POSITIONED, PROBABLY THE CAUSE FOR THE RECURRING DISLOCATION WAS INSUFFICIENT SOFT TISSUE TENSION OR SLIGHT IMMEDIATE SUBSIDENCE. NO REASON TO SUSPECT THAT A MALFUNCTIONING DEVICE CAUSED THE PROBLEM. BATCH REVIEW PERFORMED ON 09 MAY 2016. LOT 134964: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12 DECEMBER 2013. EXPIRATION DATE: 2018-10-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. MECTACER DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE L + 4, CODE 01.29.210, LOT. 152618 (K112115) (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26 OCTOBER 2015. EXPIRATION DATE: 2020-10-10. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. VERSAFITCUP CC TRIO ACETABULAR SHELL Ø 62, CODE 01.26.45.0062, LOT. 144699 (K103352) [NOT EXPLANTED] (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10 OCTOBER 2014. EXPIRATION DATE: 2019-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. NOT AVAILABLE.

Description of Event or Problem · 1

THE PATIENT HAD 2 DISLOCATIONS AFTER THE PRIMARY SURGERY, SO THE SURGEON DECIDED TO REVISE HEAD AND LINER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302891 VERSAFITCUP CC FLAT PE HC LINER Ø 36/G FLAT PE LINER LZO MEDACTA INTERNATIONAL SA 134964

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention