FDA Adverse Event Malfunction Summary report: N

METABOLIC MEASUREMENT CART

MDR report key: 56444 · Received December 6, 1996

Report

Report Number
2050001-1996-00040
Event Type
Malfunction
Date Received
December 6, 1996
Date of Event
November 13, 1996
Report Date
December 6, 1996
Manufacturer
SENSOR MEDICS CORP.
Product Code
BZC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

WHILE SETTING UP THE MODEL 2900, THE USER OBSERVED THAT THE CO2 ANALYZER ZERO AND SPAN COULD NOT BE CALIBRATED. IT WAS DETERMINED THAT A COMPONENT ON THE CO2 ANALYZER HAD BURNT. THERE WERE NO PTS INVOLVED AND THERE WERE NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 METABOLIC MEASUREMENT CART METABOLIC MEASUREMENT SYSTEM BZC SENSOR MEDICS CORP. 2900 NA

Patients

Seq Age Sex Outcome Treatment
1 NA