FDA Adverse Event Injury Summary report: N

GUNTHER TULIP FEMORAL VENA CAVA FILTER SET

MDR report key: 5640445 · Received May 9, 2016

Report

Report Number
1820334-2016-00183
Event Type
Injury
Date Received
May 9, 2016
Date of Event
February 8, 2010
Report Date
July 27, 2017
Manufacturer
COOK INC
Product Code
DTK
PMA / PMN Number
K043509
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

WILLIAM COOK EUROPE INITIALLY REPORTED EVENT UNDER MFR REPORT # 3002808486-2014-00062. NEW INFORMATION WAS RECEIVED IDENTIFYING THAT THE PRODUCT WAS A COOK INC. MANUFACTURED DEVICE. AN UPDATED EMDR REPORT WAS SUBMITTED UNDER COOK INC. MFR REPORT # 1820334-2016-00183, FOLLOW UP # 2. THIS WAS IN REFERENCE TO WILLIAM COOK EUROPE MFR REPORT # 3002808486-2014-00062. COOK IS NOW SUBMITTING AN INITIAL REPORT TO CORRECT THIS ISSUE. (B)(4). INVESTIGATION- INVESTIGATION IS BASED ON THE LIMITED INFORMATION IN THE EVENT DESCRIPTION (MAUDE REPORT MW5037141) SINCE NO CLINICAL IMAGING OR DEVICE IS RETURNED. PATIENT'S MEDICAL RECORDS ARE UNKNOWN, THEREFORE DETAILS RELATED TO FILTER IMPLANT PROCEDURE ARE UNKNOWN AND PAIN ONSET DATE UNKNOWN. UNABLE TO COMMENT ON PERFORATIONS AND UNSUCCESSFULLY RETRIEVAL ATTEMPTS AND NEED FOR A "FOURTH SURGERY" WITHOUT CLINICAL IMAGES. HOWEVER, FILTER PERFORATION OF THE VENA CAVA WALL IS A WELL KNOWN RISK IN PUBLISHED MEDICAL SCIENTIFIC LITERATURE. ALSO LITERATURE DESCRIBES DIFFERENT RETRIEVAL TECHNIQUES USED FOR DIFFICULT RETRIEVALS. A REFERENCE IS MADE TO THE POTENTIAL ADVERSE EVENTS IN THE INSTRUCTIONS FOR USE. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. BASED ON LIMITED INFORMATION UNABLE TO DETERMINE THE EXACT ROOT CAUSE FOR THIS EVENT. UPDATED ON 15MAR2016 : COMPLAINT REOPENED SINCE LOT WAS PROVIDED. NO RELEVANT NOTES ON WORK ORDER, NOR ON FILTER LOT 1654279.4. NO OTHER COMPLAINT HAS BEEN REGISTERED UNDER THIS LOT. GUNTHER TULIP FEMORAL VENA CAVA FILTER SET MANUFACTURED/INSPECTED ACCORDING TO SPECIFICATIONS. NO OTHER ADDITIONAL INFORMATION RECEIVED AT THE MOMENT, BUT IF ADDITIONAL INFORMATION IS RECEIVED, THE REPORT WILL BE REOPENED FOR FURTHER INVESTIGATION. UNTIL THEN INVESTIGATION PRECIOUSLY SUBMITTED IS STILL VALID AND REMAINS UNCHANGED.

Additional Manufacturer Narrative · 1

THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 1

INVESTIGATION- THE INVESTIGATION RESULTS PREVIOUSLY SUBMITTED ARE STILL VALID AND REMAIN UNCHANGED. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. INVESTIGATION IS BASED ON THE LIMITED INFORMATION IN THE EVENT DESCRIPTION (MAUDE REPORT MW5037141) SINCE NO CLINICAL IMAGING OR DEVICE IS RETURNED. PATIENT'S MEDICAL RECORDS ARE UNKNOWN, THEREFORE DETAILS RELATED TO FILTER IMPLANT PROCEDURE ARE UNKNOWN AND PAIN ONSET DATE UNKNOWN. UNABLE TO COMMENT ON PERFORATIONS AND UNSUCCESSFULLY RETRIEVAL ATTEMPTS AND NEED FOR A "FOURTH SURGERY" WITHOUT CLINICAL IMAGES. HOWEVER, FILTER PERFORATION OF THE VENA CAVA WALL IS A WELL KNOWN RISK IN PUBLISHED MEDICAL SCIENTIFIC LITERATURE. ALSO LITERATURE DESCRIBES DIFFERENT RETRIEVAL TECHNIQUES USED FOR DIFFICULT RETRIEVALS. A REFERENCE IS MADE TO THE POTENTIAL ADVERSE EVENTS IN THE INSTRUCTIONS FOR USE. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. BASED ON LIMITED INFORMATION UNABLE TO DETERMINE THE EXACT ROOT CAUSE FOR THIS EVENT. UPDATED ON 15MAR2016 : COMPLAINT REOPENED SINCE LOT WAS PROVIDED. NO RELEVANT NOTES ON WORK ORDER, NOR ON FILTER LOT 1654279.4. NO OTHER COMPLAINT HAS BEEN REGISTERED UNDER THIS LOT. GUNTHER TULIP FEMORAL VENA CAVA FILTER SET MANUFACTURED/INSPECTED ACCORDING TO SPECIFICATIONS.

Additional Manufacturer Narrative · 1

INITIAL MDR WAS SUBMITTED BY WILLIAM COOK EUROPE UNDER REFERENCE # 3002808486-2014-00062. ADDITIONAL INFORMATION PROVIDED ON 09MAR2016 DETERMINED THIS DEVICE WAS MANUFACTURED BY CINC. SUPPLEMENTAL MDR# 1820334-2016-00183. INVESTIGATION IS BASED ON THE LIMITED INFORMATION IN THE EVENT DESCRIPTION (MAUDE REPORT MW5037141) SINCE NO CLINICAL IMAGING OR DEVICE IS RETURNED. PATIENT'S MEDICAL RECORDS ARE UNKNOWN, THEREFORE DETAILS RELATED TO FILTER IMPLANT PROCEDURE ARE UNKNOWN AND PAIN ONSET DATE UNKNOWN. UNABLE TO COMMENT ON PERFORATIONS AND UNSUCCESSFULLY RETRIEVAL ATTEMPTS AND NEED FOR A "FOURTH SURGERY" WITHOUT CLINICAL IMAGES. HOWEVER, FILTER PERFORATION OF THE VENA CAVA WALL IS A WELL KNOWN RISK IN PUBLISHED MEDICAL SCIENTIFIC LITERATURE. ALSO LITERATURE DESCRIBES DIFFERENT RETRIEVAL TECHNIQUES USED FOR DIFFICULT RETRIEVALS. A REFERENCE IS MADE TO THE POTENTIAL ADVERSE EVENTS IN THE INSTRUCTIONS FOR USE. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. BASED ON LIMITED INFORMATION UNABLE TO DETERMINE THE EXACT ROOT CAUSE FOR THIS EVENT. UPDATED ON 15MAR2016 : COMPLAINT REOPENED SINCE LOT WAS PROVIDED. NO RELEVANT NOTES ON WORK ORDER, NOR ON FILTER LOT 1654279.4. NO OTHER COMPLAINT HAS BEEN REGISTERED UNDER THIS LOT. GUNTHER TULIP FEMORAL VENA CAVA FILTER SET MANUFACTURED/INSPECTED ACCORDING TO SPECIFICATIONS. NO OTHER ADDITIONAL INFORMATION RECEIVED AT THE MOMENT, BUT IF ADDITIONAL INFORMATION IS RECEIVED, THE REPORT WILL BE REOPENED FOR FURTHER INVESTIGATION. UNTIL THEN INVESTIGATION PRECIOUSLY SUBMITTED IS STILL VALID AND REMAINS UNCHANGED.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION IS BASED ON THE LIMITED INFORMATION IN THE EVENT DESCRIPTION (MAUDE REPORT MW5037141) SINCE NO CLINICAL IMAGING OR DEVICE IS RETURNED. PATIENT'S MEDICAL RECORDS ARE UNKNOWN, THEREFORE DETAILS RELATED TO FILTER IMPLANT PROCEDURE ARE UNKNOWN AND PAIN ONSET DATE UNKNOWN. UNABLE TO COMMENT ON PERFORATIONS AND UNSUCCESSFULLY RETRIEVAL ATTEMPTS AND NEED FOR A "FOURTH SURGERY" WITHOUT CLINICAL IMAGES. HOWEVER, FILTER PERFORATION OF THE VENA CAVA WALL IS A WELL KNOWN RISK IN PUBLISHED MEDICAL SCIENTIFIC LITERATURE. ALSO LITERATURE DESCRIBES DIFFERENT RETRIEVAL TECHNIQUES USED FOR DIFFICULT RETRIEVALS. A REFERENCE IS MADE TO THE POTENTIAL ADVERSE EVENTS IN THE INSTRUCTIONS FOR USE. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. BASED ON LIMITED INFORMATION UNABLE TO DETERMINE THE EXACT ROOT CAUSE FOR THIS EVENT. UPDATED ON 15MAR2016: COMPLAINT REOPENED SINCE LOT WAS PROVIDED. NO RELEVANT NOTES ON WORK ORDER, NOR ON FILTER LOT 1654279.4. NO OTHER COMPLAINT HAS BEEN REGISTERED UNDER THIS LOT. GUNTHER TULIP FEMORAL VENA CAVA FILTER SET MANUFACTURED/INSPECTED ACCORDING TO SPECIFICATIONS. NO OTHER ADDITIONAL INFORMATION RECEIVED AT THE MOMENT, BUT IF ADDITIONAL INFORMATION IS RECEIVED, THE REPORT WILL BE REOPENED FOR FURTHER INVESTIGATION. UNTIL THEN INVESTIGATION PRECIOUSLY SUBMITTED IS STILL VALID AND REMAINS UNCHANGED.

Additional Manufacturer Narrative · 1

(B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION.

Additional Manufacturer Narrative · 1

IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'TULIP, MIGRATION, TILT, VENA CAVA PERFORATION, DIFF REMOVE, EROSION OF VERTEBRAL BODY, PAIN'. MANIPULATION IN THE AREA OF THE FILTER IMPLANT MAY CAUSE MIGRATION OR CONTRIBUTE TO CHANGES IN THE FILTER CONFIGURATION AND PLACEMENT. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. FILTER TILT IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE AND MAY OCCUR DURING PLACEMENT OR DURING IMPLANTING PERIOD. THERE IS NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

A MALE UNKNOWN AGE UNDERWENT AN IVC FILTER PLACEMENT ON (B)(6) 2010. THE FILTER WAS SUPPOSED TO BE A RETRIEVABLE TEMPORARY FILTER. HOWEVER, THE FILTER WAS UNSUCCESSFULLY REMOVED. THE PATIENT HAS BEEN SUFFERING FROM PAIN SINCE 2010. THE FILTER HAS NOW PERFORATED THE IVC AND TOUCHING THE VERTEBRAE. THE PATIENT IS AWAITING THE MEANS OF GETTING THE DEVICE REMOVED. THE PATIENT'S MEDICAL RECORDS SHOW THE COOK GUNTHER TULIP WAS DEFECTIVE AND REQUIRES A FOURTH SURGERY TO HAVE THE DEVICE REMOVED FROM THE IVC. A FOURTH SURGERY WAS REQUIRED TO HAVE THE DEVICE REMOVED FROM THE IVC.

Description of Event or Problem · 1

AS REPORTED UNDER MW5037141: A MALE UNKNOWN AGE UNDERWENT AN IVC FILTER PLACEMENT ON (B)(6) 2010. THE FILTER WAS SUPPOSED TO BE A RETRIEVABLE TEMPORARY FILTER. HOWEVER, THE FILTER WAS UNSUCCESSFULLY REMOVED. THE PATIENT HAS BEEN SUFFERING FROM PAIN SINCE 2010. THE FILTER HAS NOW PERFORATED THE IVC AND TOUCHING THE VERTEBRAE. THE PATIENT IS AWAITING THE MEANS OF GETTING THE DEVICE REMOVED. THE PATIENT'S MEDICAL RECORDS SHOW THE COOK GUNTHER TULIP WAS DEFECTIVE AND REQUIRES A FOURTH SURGERY TO HAVE THE DEVICE REMOVED FROM THE IVC. A FOURTH SURGERY WAS REQUIRED TO HAVE THE DEVICE REMOVED FROM THE IVC. ADDITIONAL INFORMATION RECEIVED JANUARY 30, 2017: IT IS ALLEGED IN THE PENDING LAWSUIT THAT PT SUFFERS FROM MIGRATION, TILT, VENA CAVA PERFORATION, AND OTHER: EROSION INTO VERTEBRAL BODY.

Description of Event or Problem · 1

AS REPORTED UNDER MW5037141: A MALE UNKNOWN AGE UNDERWENT AN IVC FILTER PLACEMENT ON (B)(6) 2010. THE FILTER WAS SUPPOSED TO BE A RETRIEVABLE TEMPORARY FILTER. HOWEVER, THE FILTER WAS UNSUCCESSFULLY REMOVED. THE PATIENT HAS BEEN SUFFERING FROM PAIN SINCE 2010. THE FILTER HAS NOW PERFORATED THE IVC AND TOUCHING THE VERTEBRAE. THE PATIENT IS AWAITING THE MEANS OF GETTING THE DEVICE REMOVED. THE PATIENT'S MEDICAL RECORDS SHOW THE COOK GUNTHER TULIP WAS DEFECTIVE AND REQUIRES A FOURTH SURGERY TO HAVE THE DEVICE REMOVED FROM THE IVC. A FOURTH SURGERY WAS REQUIRED TO HAVE THE DEVICE REMOVED FROM THE IVC. ADDITIONAL INFORMATION RECEIVED JANUARY 30, 2017: IT IS ALLEGED IN THE PENDING LAWSUIT THAT PT SUFFERS FROM MIGRATION, TILT, VENA CAVA PERFORATION, AND OTHER: EROSION INTO VERTEBRAL BODY.

Description of Event or Problem · 1

A MALE UNKNOWN AGE UNDERWENT AN IVC FILTER PLACEMENT ON (B)(6) 2010. THE FILTER WAS SUPPOSED TO BE A RETRIEVABLE TEMPORARY FILTER. HOWEVER, THE FILTER WAS UNSUCCESSFULLY REMOVED. THE PATIENT HAS BEEN SUFFERING FROM PAIN SINCE 2010. THE FILTER HAS NOW PERFORATED THE IVC AND TOUCHING THE VERTEBRAE. THE PATIENT IS AWAITING THE MEANS OF GETTING THE DEVICE REMOVED. THE PATIENT'S MEDICAL RECORDS SHOW THE COOK GUNTHER TULIP WAS DEFECTIVE AND REQUIRES A FOURTH SURGERY TO HAVE THE DEVICE REMOVED FROM THE IVC. A FOURTH SURGERY WAS REQUIRED TO HAVE THE DEVICE REMOVED FROM THE IVC.

Description of Event or Problem · 1

A MALE UNKNOWN AGE UNDERWENT AN IVC FILTER PLACEMENT ON (B)(6) 2010. THE FILTER WAS SUPPOSED TO BE A RETRIEVABLE TEMPORARY FILTER. HOWEVER, THE FILTER WAS UNSUCCESSFULLY REMOVED. THE PATIENT HAS BEEN SUFFERING FROM PAIN SINCE 2010. THE FILTER HAS NOW PERFORATED THE IVC AND TOUCHING THE VERTEBRAE. THE PATIENT IS AWAITING THE MEANS OF GETTING THE DEVICE REMOVED. THE PATIENT'S MEDICAL RECORDS SHOW THE COOK GUNTHER TULIP WAS DEFECTIVE AND REQUIRES A FOURTH SURGERY TO HAVE THE DEVICE REMOVED FROM THE IVC. A FOURTH SURGERY WAS REQUIRED TO HAVE THE DEVICE REMOVED FROM THE IVC.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 12/02/2016 AS FOLLOWS: PLAINTIFF ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2010 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO DEEP VEIN THROMBOSIS AND PULMONARY EMBOLI. PLAINTIFF IS ALLEGING DEVICE MIGRATION, TILT, VENA CAVA PERFORATION, EROSION OF VERTEBRAL BODY, LOWER BACK PAIN, DECREASED RANGE OF MOTION, SLEEPLESSNESS. PLAINTIFF ALLEGES ATTEMPTED RETRIEVAL ON (B)(6) 2010 AND (B)(6) 2010 AND SUCCESSFUL RETRIEVAL ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
297925 GUNTHER TULIP FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC N/A

Patients

Seq Age Sex Outcome Treatment
1 28 YR Life Threatening| R