FDA Adverse Event Malfunction Summary report: N

ATTUNE FEMORAL IMPACTOR

MDR report key: 5637858 · Received May 6, 2016

Report

Report Number
1818910-2016-18777
Event Type
Malfunction
Date Received
May 6, 2016
Date of Event
April 27, 2016
Report Date
April 27, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONFIRMED THAT THE IMPACTORS HAD BROKEN EXPERT OPINION INDICATES THAT THE FAILURES ARE ASSOCIATED WITH ENVIRONMENTAL STRESS CRACKING (ESC). THE ATTUNE IMPACTORS HAVE BEEN ANNEALED TO REDUCE RESIDUAL STRESSES. LAB TESTS INDICATE SIGNIFICANT REDUCTION IN RESIDUAL STRESSES OF ANNEALED SAMPLES IN COMPARISON TO UN-ANNEALED SAMPLES HOWEVER A COMPLETE REDUCTION IN STRESSES IS NOT ACHIEVABLE. A FIELD SAFETY NOTICE WAS ISSUED IN (B)(6 2014 STATING TO REDUCE THE POSSIBILITY OF LEAVING FRAGMENTS IN PATIENTS, THE COMPANY SUGGESTS ADHERING TO THE INSTRUCTIONS FOR USE (IFU), WHICH INCLUDE INSPECTING THE IMPACTOR¿S TO ENSURE THAT NO INSTRUMENTS OR PIECES OF INSTRUMENTS ARE LEFT IN THE SURGICAL SITE PRIOR TO CLOSURE. IT SHOULD BE NOTED THAT CAPA-003835 HAS BEEN INITIATED AND CAN BE REFERENCED FOR FURTHER DETAILS. THE COMPLAINT SHALL BE CLOSED WITH A JUSTIFIED CONCLUSION; IT WILL BE ENTERED INTO THE COMPLAINT DATABASE AND MONITORED THROUGH TREND ANALYSIS. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PRODUCT SPLINTERED - ONE OF THE IMPACTOR HANDLES BROKE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
293355 ATTUNE FEMORAL IMPACTOR KNEE INSTRUMENT/TRIAL LXH DEPUY ORTHOPAEDICS, INC. AU4300981

Patients

Seq Age Sex Outcome Treatment
1