FDA Adverse Event
Other
Summary report: N
MICROSCAN DRIED GRAM POSITIVE PANELS
MDR report key: 563363
·
Received May 24, 2004
Report
- Report Number
- 2919016-2004-00001
- Event Type
- Other
- Date Received
- May 24, 2004
- Date of Event
- March 17, 2004
- Report Date
- April 22, 2004
- Manufacturer
- DADE BEHRING, INC.
- Product Code
- LON
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IN 2004, MMWR (MORBIDITY AND MORTALITY WEEKLY) A PUBLICATION OF THE CENTERS FOR DISEASE CONTROL AND PREVENTION OF THE CENTERS FOR DISEASE CONTROL AND PREVENTION (CDC) REPORTED ON ANTIMICROBIAL SUSCEPTIBILITY RESULTS OBTAINED FOR A VANCOMYCIN RESISTANT STAPHLYOCOCCUS AUREUS (VRSA) TESTED ON MICROSCAN OVERNIGHT PANELS. THE CLINICAL ISOLATE REPORTED ON MIC=4 MCG/ML ON THE MICROSCAN OVERNIGHT PANELS. THIS PROMPTED THE USER TO INITIATE CONFIRMATORY TESTING PER CDC VISA/VRSA TESTING GUIDELINES. THE ISOLATE WAS RESISTANT TO VANCOMYCIN WITH AN MIC=64 MCG/ML USING THE NCCLS BROTH MICRODILUTION METHOD. THIS IS ONLY THE THIRD VRSA ISOLATE ENCOUNTERED TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSCAN DRIED GRAM POSITIVE PANELS | ANTIMICROBIAL SUSCEPTIBILITY TEST PANEL | LON | DADE BEHRING, INC. | NA | VARIOUS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |