FDA Adverse Event Injury Summary report: N

NM MILLENNIUM MPR COLLIMATOR

MDR report key: 563044 · Received December 30, 2004

Report

Report Number
9613299-2004-00010
Event Type
Injury
Date Received
December 30, 2004
Date of Event
November 29, 2004
Report Date
December 30, 2004
Manufacturer
GE MEDICAL SYSTEMS, F.I. HAIFA
Product Code
IZX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NO INFORMATION

Narratives

Description of Event or Problem · 1

A REPORT WAS RECEIVED CONCERNING A PT NOSE INJURY IN MPR SYSTEM. THE EVENT OCCURRED IN PREPARATION FOR THYROID SCAN WITH LEGP COLLIMATOR. THE USER LOWERED THE DETECTOR HEAD AT NORMAL SPEED, USING THE HAND HELD CONTROLLER WHILE PT WAS LYING ON BED. WITHOUT BEING NOTICED BY THE OPERATOR, THE DETECTOR CAME IN CONTACT TO THE PT NOSE. THE GANTRY MOVEMENT WAS STOPPED UPON PSD ACTIVATION, BUT THE PT NOSE WAS DAMAGED; POTENTIALLY BROKEN. THE SCAN WAS STOPPED AND NO FURTHER SCANS WERE MADE TO THE PT. SYSTEM WAS RECHECKED BY FILED ENGINEERING FOR NORMAL SPD OPERATION (STOP MOTION UPON SLIGHT CONTACT BY THE OPERATOR). THE SYSTEM BEHAVED NORMALLY AND WAS RETURNED TO NORMAL USE ROUTINES FOR THE SAME TYPE OF SCANS WITH OTHER PTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NM MILLENNIUM MPR COLLIMATOR COLLIMATOR IZX GE MEDICAL SYSTEMS, F.I. HAIFA NI NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other