PRODIGY AUTOCODE
Report
- Report Number
- 3005862821-2016-00024
- Event Type
- Injury
- Date Received
- May 4, 2016
- Date of Event
- March 16, 2016
- Report Date
- March 23, 2016
- Manufacturer
- OK BIOTECH CO., LTD.
- Product Code
- NBW
- PMA / PMN Number
- K073118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
SUSPECTED DEVICE EVALUATED BY OK BIOTECH AND CALCULATED THAT THE METER OPERATED WITHIN SPECIFICATIONS. THE STANDBY CURRENT TEST IS 16.3-A. THE CRITERIA IS <55-A. PASS. METER SETTING, AUDIO AND ALL BUTTONS FUNCTION ARE OK. TESTED THE SUSPECTED METER WITH IN HOUSE CONTROL SOLUTION AND RETAIN STRIPS (STRIP LOT NUMBER: D150414-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW ARE 55/54 MG/DL; FOR LEVEL HIGH ARE 268/270 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25 APPROXIMATELY 70 MG/DL; LEVEL HIGH 210 APPROXIMATELY 310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS. TESTED PATIENT'S RETURNED STRIPS (LOT NUMBER: D150414-1). THE CONTROL SOLUTION TESTS FOR LEVEL LOW WERE 58 MG/DL; FOR LEVEL HIGH WERE 291 MG/DL. THE CONTROL SOLUTION RANGES ARE: LEVEL LOW 25 APPROXIMATELY 70 MG/DL; LEVEL HIGH 210 APPROXIMATELY 310 MG/DL. ALL RESULTS WERE WITHIN THE ACCEPTANCE RANGE. PASS.
PRODIGY DIABETES CARE RECEIVED A CALL ON (B)(6) 2016 REPORTING A MEDICAL INTERVENTION THAT OCCURED ON (B)(6) 2016 AT 1:30AM. PATIENT'S DAUGHTER (RM) CALLED IN STATING PATIENT GOT A READING WITH THE PRODIGY METER OF 28MG/DL WHICH PROMPTED HER TO TAKE PATIENT TO HOSPITAL. PATIENT TOOK LANTUS, ATE CHOCOLATE AND DRANK A SODA BEFORE BEDTIME OF THE EVENT. (RM) STATED THAT PATIENT' BLOOD GLUCOSE WAS TESTED SEVERAL ADDITIONAL TIMES BEFORE BEING TAKEN TO THE HOSPITAL WITH THE FOLLOWING RESULTS: 227, 237, 147, 227, 97, 64, 155, 137, 71 AND 48MG/DL. PATIENT'S NORMAL BLOOD GLUCOSE READING AROUND THE TIME OF THE EVENT IS BETWEEN 130-135MG/DL. (RM) STATED THAT PATIENT WAS EXPERIENCING SIGNS OF SHAKINESS. PATIENT WENT TO THE ER ON HER OWN. PATIENT'S BLOOD GLUCOSE UPON ARRIVAL AT ER WAS 204MG/DL. NO ADDITIONAL GLUCOSE TESTING WAS PERFORMED WHILE AT THE HOSPITAL AND NO TREATMENT WAS ADMINISTERED. PATIENT'S TOTAL STAY IN HOSPITAL WAS APPROXIMATELY 1/2 HOUR. PRODIGY DIABETES CARE SENT REPLACEMENT AND PREPAID ENVELOPE REQUESTING RETURN OF SUSPECT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 287709 | PRODIGY AUTOCODE | BLOOD GLUCOSE MONITORING DEVICE | NBW | OK BIOTECH CO., LTD. | 51850 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention | ANTENOLOL 25MG 1/2 PILL X1 |