FDA Adverse Event Other Summary report: N

DRAEGER MEDICAL, DISTRIBUTED BY ABBOTT LABORATORIES

MDR report key: 562565 · Received December 10, 2004

Report

Report Number
MW1033980
Event Type
Other
Date Received
December 10, 2004
Report Date
December 10, 2004
Manufacturer
DRAEGER MEDICAL
Product Code
CAD
Adverse Event
Yes
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

DRAEGER VAPOR 2000 VAPORIZORS USED W/FABIUS ANESTHESIA MACHINES, BOTH SERVOFLURANE AND ISOFLURANE: CHRONIC FAILURES, LEAKS, CAUSED BY DEGENERATED SEAL GASKETS AND JAMMED FILL PORT MECHANISMS. HAVE HAD 14 VAPORIZER FAILURES UNDER WARRANTY IN 6 MOS. FOUND DEGRADED SEAL GASKETS AND/OR JAMMED FILL PORT MECHANISMS IN UNITS EXAMINED PRIOR TO RETURNING FOR WARRANTY EXCHANGE. HAVE EXP DELAYS IN GETTING REPLACEMENTS DUE TO BACKORDERS. DRAEGER CONFIRMED THEY ARE AWARE OF BOTH PROBLEMS. JAMMED MECHANISMS DUE TO PROBLEM WITH LATCH MECHANISM - IF LATCH IS CLOSED ALL THE WAY DURING FILL PROCESS CAUSES JAMMING. GASKET PROBLEM IS DUE TO CHANGE MADE BY SUPPLIER OF GASKET. DRAEGER HAS PROVIDED NOTIFICATION OF THESE PROBLEMS ONLY THROUGH THEIR SVC NOTES AVAILABLE ONLY TO PAID SUBSCRIBERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRAEGER MEDICAL, DISTRIBUTED BY ABBOTT LABORATORIES ANESTHESIA VAPORIZER CAD DRAEGER MEDICAL VAPOR 2000 *

Patients

Seq Age Sex Outcome Treatment
1 * Other