FDA Adverse Event
Injury
Summary report: N
SERVO-I
MDR report key: 562424
·
Received December 22, 2004
Report
- Report Number
- 562424
- Event Type
- Injury
- Date Received
- December 22, 2004
- Date of Event
- November 3, 2004
- Report Date
- December 20, 2004
- Manufacturer
- MAQUET, INC.
- Product Code
- CBK
- Report Source
- User Facility report
- Reporter Location
- NY, US
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
Description of Event or Problem · 1
THE VENTILATOR BATTERY FAILED TO COMMUNICATE TO PROVIDE "LIMITED BATTERY CAPACITY" ALARM 10 MINUTES BEFORE THE END OF LIFE OF THE BATTERY. ALSO WHEN ALARMS ACTIVATED AT "NO BATTERY CAPACITY", IT LASTED FOR "30S" INSTEAD OF A MINIMUM 3 MINUTES. OVERALL IT APPEARS TO BE SOFTWARE GLITCH IN READING BATTERY LIFE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SERVO-I | LIFE SUPPORT VENTILATOR | CBK | MAQUET, INC. | SERVO-I | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention | SWITCHED TO ANOTHER VENTILATOR. |