DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2016-00288
- Event Type
- Malfunction
- Date Received
- April 28, 2016
- Date of Event
- April 10, 2016
- Report Date
- April 10, 2016
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INTUITIVE SURGICAL, INC. (ISI) HAS RECEIVED THE PART INVOLVED WITH THIS COMPLAINT. FAILURE ANALYSIS INVESTIGATION WAS ABLE TO REPLICATE THE CUSTOMER REPORTED FAILURE. THE PART WAS INSTALLED ONTO AN IN-HOUSE TEST SYSTEM AND FAILED TO POWER ON. THE PART SHOULD MEASURE AT +48 VOLTS BUT WAS FOUND TO MEASURE AT 0 VOLTS. THE POWER SUPPLY WAS FOUND DEFECTIVE. THIS COMPLAINT IS BEING REPORTED DUE TO A DA VINCI SYSTEM MALFUNCTION RENDERING THE DA VINCI SYSTEM UNAVAILABLE FOR USE AFTER THE START OF A SURGICAL PROCEDURE. ALTHOUGH NO PATIENT HARM OCCURRED, IF THIS MALFUNCTION WERE TO RECUR IT COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE PATIENT SIDE CART (PSC) WENT INTO BATTERY MODE. THE CUSTOMER TRIED TO RESOLVE THE ISSUE BY PRESSING THE EMERGENCY POWER OFF (EPO) SWITCH AND RESTARTING THE SYSTEM BUT THE ISSUE PERSISTED. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT CONFIRMED WITH THE CUSTOMER THE POWER OUTLETS WERE WORKING BUT THE SYSTEM WAS STILL RUNNING ON BATTERY MODE AS INDICATED BY THE CHARGING LED ON THE SYSTEM, WHICH REMAINED AMBER. ON (B)(6) 2016, ISI OBTAINED ADDITIONAL INFORMATION FROM THE REPORTER. THERE WAS A DELAY OF 30 MINUTES FOR THE PROCEDURE. SURGICAL PORTS WERE PLACED; HOWEVER, THERE WAS NO REPORT OF ADDITIONAL ANESTHESIA ADMINISTERED TO THE PATIENT. DUE TO THE REPORTED ISSUE, THE SURGEON DECIDED TO CONVERT THE PROCEDURE TO LAPAROSCOPY. THE PATIENT TOLERATED THE LAPAROSCOPY PROCEDURE WELL. THERE WAS NO REPORT OF PATIENT INJURY OR HARM. AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL FIELD SPECIALIST (TFS) CONDUCTED ADDITIONAL INVESTIGATION OF THE SYSTEM ON SITE. TFS FOUND THE PSC STARTED UP ON BATTERY MODE ONLY AND THE BATTERY CHARGING DOESN'T WORK. THERE WAS NO POWER ON THE DIRECT CURRENT (DC) OUTPUT OF THE INSTALLED PSC POWER SUPPLY. THE TFS REPLACED THE PSC MEDICAL GRADE POWER SUPPLY (MGPS) TO RESOLVE THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273567 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A60P9B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |