FDA Adverse Event Other Summary report: N

DIEGO CONSOLE SINUS SHAVER HANDPIECE

MDR report key: 561100 · Received January 5, 2004

Report

Report Number
561100
Event Type
Other
Date Received
January 5, 2004
Date of Event
December 30, 2003
Report Date
January 5, 2004
Manufacturer
GYRUS ENT, LLC
Product Code
HRX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IN, US

Narratives

Description of Event or Problem · 1

THE DIEGO CONSOLE HANDPIECE TO THE GYRUS-ENT EARS NOSE THROATSINUS SHAVER BECOMES UNCOMFORTABLY HOT (105 DEGREES F) DURING USE. SEVERAL HANDPIECES WERE USED AND ALL WERE FOUND TO REACH UNCOMFORTABLY HOT TEMPERATURES. NO INJURIES TO PATIENT OR STAFF OCCURRED. EXTENSIVE TESTING WAS DONE WITH THIS DEVICE, SPECIFICALLY GYRUS-ENT (800-262-3540; WWW.GYRUS-ENT.COM), MODEL 70338000. THE HANDPIECES WERE RETURNED TO THE MANUFACTURER AND THE REPLACEMENTS HAD THE SAME HEAT PROBLEM WHEN TESTED. A COMPLAINT ABOUT THE DEVICE HAS BEEN FILED WITH THE MANUFACTURER; FACILITY IS REQUESTING THE MANUFACTURER PERFORM A QUALITY ANALYSIS ON THE HANDPIECE. FACILITY'S GYRUS-ENT REP AND CLINICAL ENGINEER, FILED THE DEVICE COMPLAINT WITH THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIEGO CONSOLE SINUS SHAVER HANDPIECE SINUS SHAVER HRX GYRUS ENT, LLC 70338000 *

Patients

Seq Age Sex Outcome Treatment
1 *