ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS
Report
- Report Number
- 3002808486-2016-00238
- Event Type
- Injury
- Date Received
- April 27, 2016
- Date of Event
- April 5, 2016
- Report Date
- April 6, 2016
- Manufacturer
- WILLIAM COOK EUROPE
- Product Code
- MIH
- UDI-DI
- 10827002346595
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
(B)(4). PT INFO: UNKNOWN AS INFORMATION WAS NOT PROVIDED. SIMILAR TO DEVICE WITH 510(K) P140016. INVESTIGATION IS STILL IN PROGRESS.
(B)(4). SIMILAR TO DEVICE WITH 510(K) P140016. (B)(4). SUMMARY OF INVESTIGATIONAL FINDINGS: BASED ON THE PROVIDED INFORMATION IT WAS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE REPORTED TILT/KINK. OF THE IMAGES IT WAS POSSIBLE TO DETERMINE THE EXISTENCE OF A SECOND DEVICE; HOWEVER THE PLACEMENT OF THIS DEVICE WAS CONSTRICTED BY THE ORIGINAL GRAFT KINK. IN ADDITION, THE KINKING SHORTENED THE AVAILABLE LENGTH FOR THE SECOND DEVICE TO SEAL ALONG THE INNER AORTIC CURVATURE ALLOWING A TYPE IA ENDOLEAK TO OCCUR. NO EVIDENCE TO SUGGEST THAT THE PRODUCT WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.
DESCRIPTION OF EVENT ACCORDING TO COMPLAINANT: THE DEVICE TILTED DURING RELEASE FROM THE DELIVERY SYSTEM, DUE TO THE PRO-FORM SUTURE. PATIENT OUTCOME: ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT PROVIDED BY THE REPORTER.
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE DEVICE TILTED DURING RELEASE FROM THE DELIVERY SYSTEM, DUE TO THE PRO-FORM SUTURE. PATIENT OUTCOME: ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT PROVIDED BY THE REPORTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 266064 | ZENITH ALPHA¿ THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS | MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | WILLIAM COOK EUROPE | 10827002346595 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Life Threatening |