FDA Adverse Event Malfunction Summary report: N

ONE TOUCH ULTRA

MDR report key: 561070 · Received May 13, 2004

Report

Report Number
2939301-2004-03043
Event Type
Malfunction
Date Received
May 13, 2004
Report Date
May 11, 2004
Manufacturer
LIFESCAN, INC.
Product Code
CFR
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

A PT REPORTED BLOOD GLUCOSE RESULTS OF 249, 172, 143 AND 143 MG/DL WITH A LIFESCAN METER, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 20% AND/OR <= 20 MG/DL, THEREBY INDICATING A POTENTIAL MALFUNCTION OF THE METER IN TERMS OF PRECISION. PT HAD MENTIONED THAT A SUNDAY THEY HAD SYMPTOMS OF HYPOGLYCEMIA AND WHEN THEY TESTED THEIR BLOOD GLUCOSE METER GAVE THEM A READING OF 48MG/DL. PT THEN ATE AND DRANK SOMETHING AND FELT BETTER AFTERWARDS. NO FURTHER INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH ULTRA GLUCOSE MONITORING SYS/KIT CFR LIFESCAN, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN