EEA 25MM SINGLE-USE STAPLER
Report
- Report Number
- 2647580-2016-00169
- Event Type
- Injury
- Date Received
- April 26, 2016
- Report Date
- April 11, 2016
- Manufacturer
- COVIDIEN, FORMERLY USSC PUERTO RICO INC
- Product Code
- GDW
- PMA / PMN Number
- K062850
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ADDITIONAL ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF NEW INFORMATION BECOMES AVAILABLE. THE ASSOCIATED CASE NUMBERS ARE: (B)(4). CITATION: J GASTROINTEST SURG (2016) 20:688¿692 DOI 10.1007/S11605-016-3087-1.
ACCORDING TO A LITERATURE REPORT: INTRODUCTION: REPORTS ON OUTCOMES AFTER DOUBLE-STAPLE TECHNIQUE (DST) FOR TOTAL AND PROXIMAL GASTRECTOMY ARE LIMITED, ORIGINATING MOSTLY FROM (B)(6) CENTERS. OUR OBJECTIVE WAS TO EXAMINE ANASTOMOTIC LEAK AND STRICTURE WITH DST FOR ESOPHAGOENTERIC ANASTOMOSIS IN GASTRIC CANCER PATIENTS. METHODS: A SINGLE INSTITUTION REVIEW WAS PERFORMED FOR PATIENTS WHO UNDERWENT TOTAL/PROXIMAL GASTRECTOMY WITH DST BETWEEN 2006 AND 2015. DST WAS PERFORMED USING TRANSORAL ANVIL DELIVERY (ORVIL) WITH END-TO-END ANASTOMOSIS. CLINICAL CHARACTERISTICS AND OUTCOMES, INCLUDING ANASTOMOTIC LEAK AND STRICTURE, WERE RECORDED. RESULTS: OVERALL, DST WAS PERFORMED IN 60 PATIENTS [TOTAL GASTRECTOMY (81.7 %, N = 49/60), PROXIMAL GASTRECTOMY (10.0 %, N=6/ 60), AND COMPLETION GASTRECTOMY (8.3 %, N =5/60)]. NEOADJUVANT CHEMOTHERAPY WAS ADMINISTERED TO 21 PATIENTS (35.0 %), AND 6 PATIENTS (10.0 %) RECEIVED EXTERNAL BEAM RADIATION THERAPY PRIOR TO COMPLETION GASTRECTOMY. OPERATIVE APPROACH WAS OPEN (51.7 %, N= 31/60), LAPAROSCOPIC (43.3 %, N =26/60), OR ROBOTIC (5.0 %, N = 3/60). ANASTOMOTIC LEAK OCCURRED IN 6.7 % (N=4/ 60), WHILE STRICTURE INDEPENDENT OF LEAK WAS IDENTIFIED IN 19.0 % (N =11/58) OF PATIENTS. COMPLICATIONS OCCURRED IN 38.3 %(N =23/60) OF PATIENTS, OF WHICH 52 % WERE CLASSIFIED AS CLAVIEN-DINDO GRADES III - V COMPLICATIONS. CONCLUSION IN THE LARGEST WESTERN SERIES OF DST FOR ESOPHAGOENTERIC ANASTOMOSES IN GASTRIC CANCER SURGERY, OUR EXPERIENCE DEMONSTRATES THAT DST IS SAFE AND EFFECTIVE WITH LOW RATES OF LEAK AND STRICTURE. [...].FOUR ANASTOMOTIC LEAKS WERE REPORTED: (B)(4) [CT-GUIDED DRAINAGE AND ANTIBIOTICS], (B)(4) [STENT PLACEMENT], (B)(4) [STENT PLACEMENT] AND (B)(4) [REOPERATION]. ALTHOUGH THESE PATIENTS HAD A NEGATIVE ESOPHAGOGRAM OR UPPER ENDOSCOPY 3 DAYS POSTOPERATIVELY, ALL FOUR PATIENTS WERE EVENTUALLY DIAGNOSED WITH ANASTOMOTIC LEAK BY COMPUTED TOMOGRAPHY (CT).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264417 | EEA 25MM SINGLE-USE STAPLER | IMPLANTABLE STAPLE | GDW | COVIDIEN, FORMERLY USSC PUERTO RICO INC | EEA25 | NOT AVAILABLE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R |