CELL-DYN EMERALD CLEANER
Report
- Report Number
- 2919069-2016-00418
- Event Type
- Malfunction
- Date Received
- April 26, 2016
- Date of Event
- April 1, 2016
- Report Date
- July 5, 2016
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- K081495
- Removal / Correction Number
- 2919069-03/24/16-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THIS FOLLOW-UP REPORT IS BEING SENT TO CORECT THE SUSPECT MEDICAL DEVICE MANUFACTURE LOCATION. THE WRONG LOCATION WAS DOCUMENTED IN THE FOLLOW UP RPEORT TO 2919069-2016-00418 - 01. REPORT ABOVE HAS BEEN UPDATED TO REFLECT THE CORRECT LOCATION. NO OTHER CHANGES HAVE BEEN DOCUMENTED IN THIS REPORT. THE INVESTIGATION TO DETERMINE CAUSE IS STILL ONGOING; A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.
(B)(4). A PRODUCT RECALL LETTER WAS ISSUED TO ALL CELL-DYN EMERALD CUSTOMERS WHO HAVE RECEIVED CELL-DYN EMERALD CLEANER (LOTS 6853, 6901, 6953). THE LETTER INFORMS THE CUSTOMER OF THE ISSUE REGARDING THE POTENTIAL FOR OUT OF RANGE CONTROLS. THE LETTER INSTRUCTS THE CUSTOMER TO DISCONTINUE USE OF THE SUSPECTED LOT AND DESTROY ANY REMAINING INVENTORY. THE CUSTOMER IS INSTRUCTED THAT IF THEY DO NOT HAVE REPLACEMENT MATERIAL AVAILABLE, THEY CAN CONTINUE TO USE THE LOT UNTIL THEY RECEIVE REPLACEMENT MATERIAL AS LONG AS THEIR CONTROLS ARE IN RANGE. THE CUSTOMER IS INSTRUCTED TO PERFORM THE DECONTAMINATION PROCEDURE PER THE OPERATOR'S MANUAL. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.
FA24MAR2016 HAS BEEN EXPANDED TO INCLUDE CELL-DYN EMERALD CLEANER LOT NUMBERS 7044, 7082, 7110 AND 7119. THE CAUSE FOR THE QC MATERIAL OUT-OF-RANGE LOW ISSUE HAS BEEN TRACED TO THE MANUFACTURING PROCESS FOR RAW MATERIAL USED IN THE CELL-DYN EMERALD CLEANER.
FA24MAR2016 HAS BEEN EXPANDED TO INCLUDE CELL-DYN EMERALD CLEANER LOT NUMBERS 6991, 7024 AND 7027. THE INVESTIGATION TO DETERMINE CAUSE IS STILL ONGOING; A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.
THE CUSTOMER WAS EXPERIENCING PLT CONTROLS OUT OF RANGE LOW WITH THE USE OF CELL-DYN EMERALD CLEANER REAGENT LOT NUMBER 6901. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265488 | CELL-DYN EMERALD CLEANER | CELL-DYN EMERALD | GKZ | ABBOTT DIAGNOSTICS DIVISION | 6901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |