FDA Adverse Event Malfunction Summary report: N

CELL-DYN EMERALD CLEANER

MDR report key: 5610017 · Received April 26, 2016

Report

Report Number
2919069-2016-00418
Event Type
Malfunction
Date Received
April 26, 2016
Date of Event
April 1, 2016
Report Date
July 5, 2016
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
PMA / PMN Number
K081495
Removal / Correction Number
2919069-03/24/16-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SENT TO CORECT THE SUSPECT MEDICAL DEVICE MANUFACTURE LOCATION. THE WRONG LOCATION WAS DOCUMENTED IN THE FOLLOW UP RPEORT TO 2919069-2016-00418 - 01. REPORT ABOVE HAS BEEN UPDATED TO REFLECT THE CORRECT LOCATION. NO OTHER CHANGES HAVE BEEN DOCUMENTED IN THIS REPORT. THE INVESTIGATION TO DETERMINE CAUSE IS STILL ONGOING; A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). A PRODUCT RECALL LETTER WAS ISSUED TO ALL CELL-DYN EMERALD CUSTOMERS WHO HAVE RECEIVED CELL-DYN EMERALD CLEANER (LOTS 6853, 6901, 6953). THE LETTER INFORMS THE CUSTOMER OF THE ISSUE REGARDING THE POTENTIAL FOR OUT OF RANGE CONTROLS. THE LETTER INSTRUCTS THE CUSTOMER TO DISCONTINUE USE OF THE SUSPECTED LOT AND DESTROY ANY REMAINING INVENTORY. THE CUSTOMER IS INSTRUCTED THAT IF THEY DO NOT HAVE REPLACEMENT MATERIAL AVAILABLE, THEY CAN CONTINUE TO USE THE LOT UNTIL THEY RECEIVE REPLACEMENT MATERIAL AS LONG AS THEIR CONTROLS ARE IN RANGE. THE CUSTOMER IS INSTRUCTED TO PERFORM THE DECONTAMINATION PROCEDURE PER THE OPERATOR'S MANUAL. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.

Additional Manufacturer Narrative · 1

FA24MAR2016 HAS BEEN EXPANDED TO INCLUDE CELL-DYN EMERALD CLEANER LOT NUMBERS 7044, 7082, 7110 AND 7119. THE CAUSE FOR THE QC MATERIAL OUT-OF-RANGE LOW ISSUE HAS BEEN TRACED TO THE MANUFACTURING PROCESS FOR RAW MATERIAL USED IN THE CELL-DYN EMERALD CLEANER.

Additional Manufacturer Narrative · 1

FA24MAR2016 HAS BEEN EXPANDED TO INCLUDE CELL-DYN EMERALD CLEANER LOT NUMBERS 6991, 7024 AND 7027. THE INVESTIGATION TO DETERMINE CAUSE IS STILL ONGOING; A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN CAUSE IS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER WAS EXPERIENCING PLT CONTROLS OUT OF RANGE LOW WITH THE USE OF CELL-DYN EMERALD CLEANER REAGENT LOT NUMBER 6901. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
265488 CELL-DYN EMERALD CLEANER CELL-DYN EMERALD GKZ ABBOTT DIAGNOSTICS DIVISION 6901

Patients

Seq Age Sex Outcome Treatment
1