ACUSON SC2000 ULTRASOUND SYSTEM
Report
- Report Number
- 3009498591-2015-00165
- Event Type
- Malfunction
- Date Received
- July 1, 2015
- Date of Event
- April 12, 2015
- Report Date
- June 15, 2018
- Manufacturer
- SIEMENS MEDICAL SOLUTIONS USA INC.
- Product Code
- IYN
- PMA / PMN Number
- K113179
- Removal / Correction Number
- 300949859110072015009C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
WHEN PATIENT NAME, PATIENT ID, PATIENT DATE OF BIRTH, AND THE PATIENT GENDER MATCH, THE SYSTEM WORKS AS EXPECTED. IN THE CASE OF THE PATIENT'S NAME HAVING A DIFFERENCE IN UPPER/LOWER CASE FROM THE ORIGINAL GENERATION OF THIS PATIENT, THE SW IS NOT CORRECTLY ACCOUNTING FOR THIS DIFFERENCE. THE SYSTEM SEES THIS AS TWO PATIENTS WITH 3 OR 4 IDENTIFIERS MATCHING, AND DOES NOT ALLOW THE CAPTURE OF CLIPS/IMAGES. THIS WILL BE RESOLVED VIA SOFTWARE RELEASE.
(B)(4). THE DEVICE REFERENCED IN THIS REPORT WAS NOT RETURNED TO SIEMENS FOR EVALUATION; HOWEVER, THE SYSTEM LOG FILES WERE CAPTURED AND ANALYZED. DURING THE ANALYSIS, IT WAS FOUND THAT THE SOFTWARE WAS NOT STORING IMAGES DUE TO "CHECK-IN" FAILURES. WHEN SUBMITTING PATIENT INFORMATION IDENTICAL TO AN EXISTING PATIENT, EXCEPT FOR CASE DIFFERENCES IN THE PATIENT NAME, A SOFTWARE BUG WILL NOT ALLOW STORING OF IMAGES. THE IMAGE STORING ISSUE WAS RESOLVED IN SOFTWARE UPDATES VA35E (RELEASED 7 OCTOBER 2015), VB10D (RELEASED 10 JUNE 2016), AND VB20A (RELEASED 17 APRIL 2017). REFERENCE COMPLAINT # (B)(4).
CUSTOMER REPORTED THAT THE PATIENT FILE WAS NOT SAVED CORRECTLY WHEN ENDING THE EXAM. THE INVESTIGATION REVEALED THAT IMAGES ARE NOT ACQUIRED DURING THE STUDY.
IT WAS REPORTED THAT A SYSTEM ERROR OCCURRED DURING AN EXAM. THERE WAS NO PATIENT ADVERSE EVENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.