FDA Adverse Event Other Summary report: N

BD LOGIC BLOOD GLUCOSE MONITOR

MDR report key: 560477 · Received May 13, 2004

Report

Report Number
2243072-2004-00225
Event Type
Other
Date Received
May 13, 2004
Date of Event
April 13, 2004
Report Date
May 12, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT RECEIVING READINGS SOMETIMES 100 POINTS HIGHER THAN THEIR OTHER METER. UNSURE WHICH IS CORRECT, BUT FEELS THE OTHER METER IS MORE ACCURATE TO THE WAY THEY FEEL. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (143, 130, 130) AS REF. POINT VS. BD READING (289, 269, 239) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 3064282

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other