BD LOGIC BLOOD GLUCOSE MONITOR
MDR report key: 560477
·
Received May 13, 2004
Report
- Report Number
- 2243072-2004-00225
- Event Type
- Other
- Date Received
- May 13, 2004
- Date of Event
- April 13, 2004
- Report Date
- May 12, 2004
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT RECEIVING READINGS SOMETIMES 100 POINTS HIGHER THAN THEIR OTHER METER. UNSURE WHICH IS CORRECT, BUT FEELS THE OTHER METER IS MORE ACCURATE TO THE WAY THEY FEEL. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (143, 130, 130) AS REF. POINT VS. BD READING (289, 269, 239) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 3064282 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |