HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT
Report
- Report Number
- 1417411-2016-00046
- Event Type
- Death
- Date Received
- April 20, 2016
- Date of Event
- March 24, 2016
- Report Date
- March 24, 2016
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- OGG
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION. IT WAS RECEIVED WITHOUT WATER IN THE BOTTLE. A VISUAL EXAM WAS PERFORMED AND IT WAS FOUND THAT THERE WAS DAMAGE ON THE THREAD OF THE ADAPTOR. FUNCTIONAL TESTING WAS PERFORMED AND THE SAMPLE WAS ATTACHED TO A HUDSON RCI FLOW METER AND INSPECTED FOR OXYGEN FLOW AT BOTH 6 L/MIN AND 10 L/ MIN USING A TSI 4000 SERIES FLOWMETER. IT WAS FOUND TO BE DIFFICULT TO CORRECTLY THREAD THE ADAPTOR TO THE FLOW METER TO FORM A PROPER SEAL (I.E. IMPROPER SCREW THREADING); HOWEVER, ONCE THE AQUAPAK WAS THREADED PROPERLY ONTO THE FLOW METER, AIRFLOW WAS CONFIRMED, WITH 5.97 L/MIN (WITH FLOW METER SET TO 6 L/MIN) AND 9.8-9.9 L/MIN (WITH FLOW METER SET TO 10 L/MIN), RESPECTIVELY. IN ADDITION, THE SAMPLE WAS TESTED FOR LEAKAGE USING SAWGELOK SNOOP LIQUID LEAK DETECTION SOAP AND IT WAS OBSERVED THAT BUBBLES WERE ESCAPING FROM THE BOTTLE ASSEMBLY. ADDITIONAL TESTING WAS PERFORMED AND THE ADAPTOR WAS ATTACHED TO AN UNOPENED 340ML AQUAPAK PRE-FILLED HUMIDIFIER BOTTLES. AIRFLOW WAS CONFIRMED WITH 6.35 L/MIN (WITH FLOW METER SET TO 6.5 L/MIN) AND 10.55 L/MIN (WITH FLOW METER SET TO 10 L/MIN), RESPECTIVELY. OTHER REMARKS: BASED ON THE INVESTIGATION PERFORMED THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. THE SAMPLE PASSED THE FUNCTIONAL TESTING AND THE COMPLAINT COULD NOT BE REPLICATED UNDER IFU/LABEL INSTRUCTIONS. THE SAMPLE PASSED BOTH WITH AND WITHOUT WATER IN THE BOTTLE. REGARDING THE DAMAGE ON THE THREAD OF THE ADAPTOR, A CAPA WAS OPENED TO ADDRESS THIS ISSUE. ALL MANUFACTURING PERSONNEL FROM THE MOLDING AREA RELATED TO THIS PROCESS WERE NOTIFIED OF THIS CUSTOMER COMPLAINT.
(B)(4). NOTE: ADDITIONAL INFORMATION REGARDING THIS EVENT AND CAUSE OF PATIENT DEATH HAS BEEN REQUESTED. TO DATE, NO ADDITIONAL INFORMATION IS AVAILABLE FOR THIS REPORT. THE DEVICE SAMPLE WAS RECEIVED, BY THE MANUFACTURER, BUT THE RESULTS OF THE EVALUATION ARE INCOMPLETE AT THE TIME OF THIS REPORT. DEVICE HISTORY RECORD REVIEW SHOWED THAT THERE WERE NO FUNCTIONAL ISSUES RELATED TO ON THE MOLDED COMPONENTS INVOLVED IN THIS COMPLAINT. CUSTOMER COMPLAINT CANNOT BE CONFIRMED BASED ONLY ON THE INFORMATION PROVIDED (THE OXYGEN DIDN'T PASS THROUGH THE AQUAPAK, TO PERFORM AN INVESTIGATION AND DETERMINE THE SOURCE OF DEFECT REPORTED IS NECESSARY TO EVALUATE THE SAMPLE INVOLVED IN THIS COMPLAINT.
THE CUSTOMER ALLEGES THAT A MALE PATIENT WITH A TRACHEOCANULA GOT INTO DISTRESS (NOT IN THE ICU) AND IT WAS DECIDED BY THE NURSE AND DOCTOR TO DELIVER OXYGEN TO THE PATIENT TO COUNTER DESATURATION. AN AQUAPAK WAS ALREADY MOUNTED ON THE FLOW METER FOR THE DELIVERY OF OXYGEN VIA THE AQUAPAK, BUT THE PATIENT DIDN'T RECOVER. THE MEDICAL STAFF CHECKED ALL PARAMETERS (TRACHEOCANULA) AND NOTICED THAT THE OXYGEN DIDN'T PASS THROUGH THE AQUAPAK. A SECOND FLOW METER WAS ALSO TESTED AND OPENED THE OXYGEN AND ALREADY FELT OXYGEN BEFORE REACHING THE BOTTLE. THIS BOTTLE REMAINED UNOPENED. THE PATIENT GOT MORE AND MORE IN DISTRESS AND EVENTUALLY DIED.
THE CUSTOMER ALLEGES THAT A MALE PATIENT WITH A TRACHEOCANULA GOT INTO DISTRESS (NOT IN THE ICU) AND IT WAS DECIDED BY THE NURSE AND DOCTOR TO DELIVER OXYGEN TO THE PATIENT TO COUNTER DESATURATION. AN AQUAPAK WAS ALREADY MOUNTED ON THE FLOW METER FOR THE DELIVERY OF OXYGEN VIA THE AQUAPAK, BUT THE PATIENT DIDN'T RECOVER. THE MEDICAL STAFF CHECKED ALL PARAMETERS (TRACHEOCANULA) AND NOTICED THAT THE OXYGEN DIDN'T PASS THROUGH THE AQUAPAK. A SECOND FLOW METER WAS ALSO TESTED AND OPENED THE OXYGEN AND ALREADY FELT OXYGEN BEFORE REACHING THE BOTTLE. THIS BOTTLE REMAINED UNOPENED. THE PATIENT GOT MORE AND MORE IN DISTRESS AND EVENTUALLY DIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242910 | HUDSON AQUAPAK 340 SW,340 ML W/040 ADAPTOR,INT | HUMIDIFIER NEBULIZER KIT | OGG | TELEFLEX MEDICAL | 492157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Death |