ECARECOORDINATOR
Report
- Report Number
- 1125873-2016-00028
- Event Type
- Malfunction
- Date Received
- April 20, 2016
- Date of Event
- March 22, 2016
- Report Date
- April 20, 2016
- Manufacturer
- PHILIPS VISICU
- Product Code
- DRG
- PMA / PMN Number
- K141706
- Removal / Correction Number
- 1125873-04-01-2016-009-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NURSE
Narratives
ENGINEERING INVESTIGATED THIS ISSUE THROUGH CODE REVIEW AND PATIENT DATA; AND DETERMINED THAT THE ISSUE WITH THE SPO2 MEASUREMENT IS ASSOCIATED WITH A SOFTWARE DEFECT. WHEN THE SOFTWARE DEFECT OCCURS, THE PATIENT'S SUBMITTED MEASUREMENTS ARE NOT POPULATED IN THE PATIENT'S DATA TABLE. THE CLINICIAN IS UNABLE TO VIEW THE PATIENT'S MEASUREMENTS IN THE ECARECOORDINATOR CLINICAL USER INTERFACE. THE CUSTOMER DOES RECEIVE ADHERENCE FLAGS INDICATING THE PATIENT'S MEASUREMENTS HAVE NOT BEEN RECEIVED. A CORRECTION TO THE SOFTWARE DEFECT HAS BEEN DEVELOPED, REVIEWED AND APPROVED. THE CORRECTION IS RELEASED AND PHILIPS VISICU IS WORKING WITH CUSTOMERS TO IMPLEMENT.
CUSTOMER REPORTED THAT THEY WERE UNABLE TO VIEW THE PATIENT'S MEASUREMENTS IN THE CLINICAL USER INTERFACE OF THE DEVICE ON THE DATE OF THE INCIDENT. HOWEVER, THE CUSTOMER COULD VIEW THE PATIENT'S MEASUREMENTS WITH A BACK END TOOL. THE CUSTOMER DID NOT REPORT ANY CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245355 | ECARECOORDINATOR | ECARECOORDINATOR | DRG | PHILIPS VISICU | 45356456091 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |