8F TRIOX SVO2/CCO PA CATHETERS
Report
- Report Number
- 2025816-2016-00024
- Event Type
- Malfunction
- Date Received
- April 20, 2016
- Report Date
- February 17, 2016
- Manufacturer
- ICU MEDICAL, INC.
- Product Code
- DQE
- PMA / PMN Number
- K06299
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
INITIAL MFG. LOT BUILD REVIEWS: A REVIEW OF THE MFG. LOT BUILD DATABASE FOR THE THREE REPORTED LOT# RECORDED LOT# 3124651 (MFG.10/2015) SHOWED (B)(4) UNITS; LOT# 3086248 (MFG.07/2015) ALSO SHOWED (B)(4) UNITS AND LOT# 3129155 (MFG. 10/2015) SHOWED (B)(4) UNITS WERE ALL MFG., TESTED, INSPECTED AND RELEASED. THERE WERE NO EXCEPTION DOCUMENTS GENERATED DURING THE LOT BUILDS. QE INVESTIGATION IS IN PROGRESS.
COMPLAINT RECEIVED REPORTING INTERMITTENT SVO2 READING ISSUES WITH USE OF 52511-14 8F TRIOX SVO2/CCO PA CATHETERS, MULTIPLE LOT# . THE INITIAL INFORMATION RECEIVED REPORTS OVER THE PAST TWO WEEKS UNSPECIFIED NUMBER OF OCCURRENCES (PRE AND POST OP) WHERE SVO2 READING FAILURES OCCURRED. THE INITIAL INFORMATION RECEIVED REPORTS ".. PRE-INSERTION CALIBRATION WOULD NOT COMPLETE EVEN WITH MULTIPLE OP MODS. ONCE FLOATED, CATHETER WOULD NOT READ SV02. UNABLE TO SEE LIGHT ON Q2+. OP MOD HAD LIGHT COMING FROM TIP.". ONE EVENT WAS DESCRIBED AS FOLLOWS " ..AFTER PLACEMENT OF CATHETER ATTEMPTS TO OBTAIN SV02 READINGS WERE UNSUCCESSFUL, THE CATHETER WAS REMOVED AND REPLACED WITH A SECOND CATHETER WHICH ALSO FAILED TO PROVIDE SVO2 READINGS. ATTENDING MEDICAL STAFF/CLINICIANS ARE HAVING TO DRAW MANUAL GASSES EVERY FOUR HOURS FOR UP TO 2 DAYS WHICH IS CAUSING "CRITICAL DELAYS IN ASSESSMENTS ... TIMELY ADMINISTRATION OF MEDICAL TREATMENTS." THERE WERE NO REPORTED PATIENT INJURIES AND OR ADVERSE PATIENT OUTCOMES. REPORTED LOT# 3124651 (MFG.10/2015) (B)(4) UNITS; 3086248 (MFG.07/2015) (B)(4) UNITS; 3129155 (MFG. 10/2015) (B)(4) UNITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244154 | 8F TRIOX SVO2/CCO PA CATHETERS | 8F TRIOX¿ SVO2/CCO PA CATHETERS | DQE | ICU MEDICAL, INC. | 52511-14 | 3124651;3086248;3129155 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |