AIRVO2 HUMIDIFIER
Report
- Report Number
- 9611451-2016-00160
- Event Type
- Malfunction
- Date Received
- April 20, 2016
- Date of Event
- March 21, 2016
- Report Date
- March 21, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K131895
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE COMPLAINT AIRVO HUMIDIFIER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED AND ELECTRICALLY TESTED. RESULTS: DURING TESTING THE AIRVO TURNED ON AND FUNCTIONED, HOWEVER NO AUDIBLE ALARM WAS HEARD. THE FAULT WAS TRACED TO A FAULTY SPEAKER AND ELECTRICAL RESISTANCE TESTING SHOWED THAT THE SPEAKER'S RESISTANCE WAS OPEN CIRCUIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 150309. CONCLUSION: THE SUPPLIER OF THE SPEAKER UNIT WAS NOTIFIED AND THEY HAVE CARRIED OUT AN INVESTIGATION. THE PROBLEM HAS BEEN TRACED TO AN ISSUE WITH THE GLUING PROCESS. THE SUPPLIER HAS TAKEN STEPS TO ENSURE THAT EACH SPEAKER IS CHECKED FOLLOWING THE GLUING PROCESS AND ANY FOUND FAULTY ARE DISCARDED. ADDITIONAL CHECKS HAVE ALSO BEEN IMPLEMENTED DURING PRODUCTION AT OUR FACILITY TO ENSURE THE SPEAKER IS WORKING AT THE TIME OF MANUFACTURE OF THE AIRVO. THE AIRVO USER MANUAL STATES THAT THE "AIRVO IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH-FLOW, WARMED AND HUMIDIFIED RESPIRATORY GASES." AND THAT " THE UNIT IS NOT INTENDED FOR LIFE SUPPORT." THE USER MANUAL ALSO CONTAINS INSTRUCTIONS ON HOW TO CHECK THE ALARM SYSTEM FUNCTIONALITY AND STATES THAT "IF EITHER ALARM SIGNAL IS ABSENT, DO NOT USE THE UNIT. CONTACT YOUR FISHER & PAYKEL HEALTHCARE REPRESENTATIVE."
A HOSPITAL IN (B)(6) REPORTED THAT THE AUDIBLE ALARM OF AN AIRVO 2 HUMIDIFIER WAS NOT FUNCTIONING. THIS WAS FOUND BEFORE PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245420 | AIRVO2 HUMIDIFIER | AIRVO2 HUMIDIFIER | BTT | FISHER & PAYKEL HEALTHCARE LTD | PT101 | 150901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |