NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2016-00162
- Event Type
- Malfunction
- Date Received
- April 20, 2016
- Report Date
- March 21, 2016
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE MANOMETER OF THE COMPLAINT RD900AFU NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR PHYSICAL DAMAGE. THE MANOMETER WAS PERFORMANCE TESTED BY FITTING INTO A KNOWN GOOD VALVE SYSTEM AND TESTED BASED ON THE NEOPUFF TECHNICAL MANUAL. RESULTS: NO PHYSICAL DAMAGE WAS OBSERVED TO THE RETURNED MANOMETER. THERE WAS ALSO NO BLOCKAGE OBSERVED ON THE MANOMETER INLET PORT. HOWEVER, THE MANOMETER FAILED THE PERFORMANCE TEST SPECIFIED IN THE NEOPUFF TECHNICAL MANUAL. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 140702. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. ALL NEOPUFF UNITS, INCLUDING THE MANOMETERS, ARE VISUALLY INSPECTED FOR ANY DEFECTS AND PERFORMANCE TESTED FOR FUNCTIONALITY PRIOR TO LEAVING THE PRODUCTION LINE. THOSE THAT FAIL ARE REJECTED. THE DAMAGE TO THE MANOMETER IS MOST LIKELY CAUSED BY SIGNIFICANT IMPACT. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF / PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT". IN ADDITION THE NEOPUFF USER INSTRUCTIONS STATE THAT THE USER SHOULD "CHECK MANOMETER READS ZERO WITH NO GAS FLOW" AND CHECK THE PRESSURE SETTINGS "PRIOR TO EVERY USE OF THE NEOPUFF". THE DEFECTIVE MANOMETER WAS REPLACED AND THE SUBJECT NEOPUFF UNIT WAS RETURNED TO THE HOSPITAL SERVICE AFTER PASSING THE PERFORMANCE CHECKS SPECIFIED IN THE NEOPUFF TECHNICAL MANUAL.
(B)(4). THE COMPLAINT DEVICE WAS RECENTLY RETURNED TO OUR SERVICE CENTRE IN (B)(4) FOR EVALUATION. OUR INVESTIGATION IS CURRENTLY IN PROGRESS AND WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED THAT THE MANOMETER ON AN RD900 NEOPUFF INFANT RESUSCITATOR WAS BLOCKED. THEY FURTHER REQUESTED A REPAIR OF THE DEVICE. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT THE MANOMETER ON AN RD900 NEOPUFF INFANT RESUSCITATOR WAS BLOCKED. THEY FURTHER REQUESTED A REPAIR OF THE DEVICE. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 246036 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 140702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |