PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2016-02400
- Event Type
- Injury
- Date Received
- April 14, 2016
- Date of Event
- March 21, 2016
- Report Date
- May 4, 2016
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND THE REPORTED NEEDLE TO CUFF MISS COULD NOT BE CONFIRMED AS ALL COMPONENTS WERE NOT RETURNED FOR ANALYSIS. A CONCLUSIVE CAUSE FOR THE REPORTED NEEDLE TO CUFF MISS/SUTURE RETRIEVAL ISSUE COULD NOT BE DETERMINED. THE TREATMENT APPEARS TO BE RELATED TO PROCEDURE CIRCUMSTANCE. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE. EIGHT UNUSED, STERILE PROGLIDE DEVICES WERE RETURNED FOR EVALUATION, TWO WITH LOT # 5121741 AND FIVE WITH LOT # 50909K2. VISUAL AND FUNCTIONAL TESTING WAS SUCCESSFULLY PERFORMED WITH NO MALFUNCTION NOTED.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER SIX PROGLIDE DEVICES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING PROGLIDE DEVICES WITH A 6F SHEATH AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, CUFF MISSES OCCURRED WITH SEVEN PROGLIDE DEVICES. AN EIGHTH PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233761 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 50909K2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |