VIDAS® TROPONIN I ULTRA
Report
- Report Number
- 3002769706-2016-00066
- Event Type
- Malfunction
- Date Received
- April 13, 2016
- Report Date
- June 14, 2017
- Manufacturer
- BIOMERIEUX SA
- Product Code
- MMI
- PMA / PMN Number
- K063243
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A CUSTOMER REPORTED A DISCREPANT RESULT WHEN USING THE VIDAS® TROPONIN TEST KIT (REF 30448). THE CUSTOMER OBTAINED A NEGATIVE RESULT ON THE VIDAS® INSTRUMENT WITH THE TROPONIN KIT WHICH LED TO THE DISCREPANT RESULT. AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS PERFORMED.
AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS PERFORMED WITH RESULTS AS FOLLOWS: THREE (3) PROBLEMS WERE CONFIRMED DURING THE COMPLAINT INVESTIGATION. A CODING ERROR BETWEEN TWO VIDAS® TECHNIQUES, UNIT ERROR; RESULTS OBTAINED WITH VIDAS TROPONIN I (LESS THAN 0.01 UG/L OR 10 NG/L) ARE LOWER THAN RESULTS OBTAINED WITH OTHER TECHNIQUES INCLUDING VIDAS HSTNI (REF 415386). THE SOLUTIONS TO THESE PROBLEMS ARE AS FOLLOWS: CODING ERROR; AS RECOMMENDED BY PROBIOQUAL THROUGH MAILING AND REPORTS, THE VIDAS USERS HAVE TO USE THE SORTING CODE XX FOR VIDAS TNIU (REF 30448) AND US FOR VIDAS HSTNI (REF 415386). UNIT ERROR; REMINDER INDICATED IN PROBIOQUAL REPORT : AS HIGH SENSITIVE TECHNIQUES ARE MORE AND MORE REPRESENTED, WE NOW USE NG/L AS UNIT. NG/L IS THE UNIT RECOMMENDED BY SCIENTISTS SOCIETIES OF CARDIOLOGY. DESPITE THE PROBIOQUAL RECOMMENDATIONS, SOME VIDAS USERS ENTERED THEIR RESULTS IN UG/L INSTEAD OF NG/L FOR VIDAS TNIU (REF 30448). QP LABORATORY ANTICIPATED THIS POTENTIAL PROBLEMS OF UNITS AND ASKED PBQ TO ADD A SCROLLING MENU TO CHOOSE UNIT WHEN ENTERING RESULTS AT THE BEGINNING OF 2016 (APPENDIX 3) RESULTS OBTAINED WITH VIDAS TROPONIN I; QP ASKED PROBIOQUAL FOR PRECISIONS ABOUT SAMPLE COMPOSITION AND EXPECTED VALUES (APPENDIX 1) : ANSWER FROM PROBIOQUAL: THE SAMPLE IS A HUMAN MATRIX LOADED WITH RECOMBINANT TROPONIN I. THERE ARE NO EXPECTED VALUES FOR BIOMÉRIEUX TECHNIQUE. THIS LEVEL OF CONTROL IS NOT ADAPTED TO VIDAS TNIU (REF 30448) BUT IS ADAPTED TO VIDAS HSTNI (REF 415386). WE KNOW THAT ARTIFICIAL SAMPLES HAVE A DIFFERENT BEHAVIOR FROM NATURAL SAMPLES. MORE THAN A DIFFERENCE OF CORRELATION BETWEEN TECHNIQUES, ARTIFICIAL AND NATURAL SAMPLES ARE NOT ANALYZED IN THE SAME WAY. VIDAS TNIU (REF 30448) AND VIDAS HSTNI (REF 415386) ARE WITHIN EXPECTED SPECIFICATIONS.
A CUSTOMER REPORTED A DISCREPANT RESULT WHEN USING THE VIDAS TROPONIN TEST KIT (REF 30448). THE CUSTOMER OBTAINED A NEGATIVE RESULT ON THE VIDAS INSTRUMENT WITH THE TROPONIN KIT WHICH LED TO THE DISCREPANT RESULT. TESTING THROUGH OTHER METHODS OBTAINED A POSITIVE RESULT. AN INVESTIGATION INTO THIS EVENT HAS BEEN INITIATED BY BIOMERIEUX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 229781 | VIDAS® TROPONIN I ULTRA | VIDAS® TROPONIN I ULTRA | MMI | BIOMERIEUX SA | 1004025610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |