ADVIA AUTOSLIDE SYSTEM

MDR report key: 5567467 · Received April 11, 2016

Report

Report Number
2432235-2016-00177
Event Type
Malfunction
Date Received
April 11, 2016
Date of Event
March 17, 2016
Report Date
October 6, 2016
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
KPA
PMA / PMN Number
K051693
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS INC. IS INVESTIGATING THE CAUSE OF THE TWO GLASS SLIDES BEING MISLABELED ON THE ADVIA AUTOSLIDE SYSTEM.

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS INC. FILED THE INITIAL MDR (2432235-2016-00177) ON APRIL 11, 2016. SEPTEMBER 14, 2016, ADDITIONAL INFORMATION: SIEMENS HEALTHCARE DIAGNOSTICS INC. (SIEMENS) HAS INVESTIGATED THE MISLABELING OF TWO GLASS SLIDES ON THE ADVIA AUTOSLIDE SYSTEM AND HAS DETERMINED THAT THERE ARE NO ISSUES WITH THE PRINTER CONTROL BOARD. SIEMENS HAS DETERMINED THAT THIS IS AN ISOLATED INCIDENT AT THIS CUSTOMER SITE AND IS RECOMMENDING THAT THE ADVIA AUTOSLIDE SYSTEM BE REPLACED AT THIS ACCOUNT DUE TO THE AGE OF THE SYSTEM AND ITS HEAVY USE. THE CAUSE OF THE EVENT IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED.

Description of Event or Problem · 1

TWO MISLABELED GLASS SLIDES WERE PRODUCED ON THE ADVIA AUTOSLIDE SYSTEM. THE FIRST GLASS SLIDE WAS LABELED WITH THE CORRECT PATIENT NAME BUT WITH THE PATIENT SAMPLE IDENTIFICATION (SID) NUMBER FOR THE SUBSEQUENT GLASS SLIDE. THE SUBSEQUENT GLASS SLIDE HAD NO PATIENT NAME AND NO SID NUMBER. NO INCORRECT RESULTS WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE TWO GLASS SLIDES BEING MISLABELED BY THE ADVIA AUTOSLIDE SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
222155 ADVIA AUTOSLIDE SYSTEM ADVIA AUTOSLIDE SYSTEM KPA SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA AUTOSLIDE SYSTEM

Patients

Seq Age Sex Outcome Treatment
1