FDA Adverse Event Malfunction Summary report: N

PFC*CALIBRATED PAT CUT GDE

MDR report key: 5563645 · Received April 8, 2016

Report

Report Number
1818910-2016-16978
Event Type
Malfunction
Date Received
April 8, 2016
Date of Event
April 4, 2016
Report Date
April 4, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
HWT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THE REPORTED EVENT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND ADDITIONAL REPORTS OF SAW CAPTURE BREAKAGE AGAINST PRODUCT CODE 865034. THE PREVIOUS REPORTS WERE INVESTIGATED AND THE ROOT CAUSES ATTRIBUTED TO PRODUCT WEAR OUT THORUGH NORMAL USE AND SERVICING. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE CURRENT REPORTED EVENT WITHOUT THE INSTRUMENT TO EXAMINE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, A NEED FOR CORRECTIVE ACTION IS NOT INDICATED DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

CAPTURE GUIDE BROKE OFF DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
217572 PFC*CALIBRATED PAT CUT GDE KNEE INSTRUMENT/TRIAL HWT DEPUY ORTHOPAEDICS, INC. 1818910 J0710

Patients

Seq Age Sex Outcome Treatment
1