FDA Adverse Event
Malfunction
Summary report: N
OT VERIO IQ METER
MDR report key: 5562391
·
Received April 8, 2016
Report
- Report Number
- 2939301-2016-16933
- Event Type
- Malfunction
- Date Received
- April 8, 2016
- Report Date
- March 16, 2016
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
ON (B)(6) 2016, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE RESULTS OF 214, 114 AND 164MG/DL OBTAINED USING THE SUBJECT DEVICE WITHIN 20 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 218777 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3824849 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |