FDA Adverse Event Other Summary report: N

BOVIE BURN MACHINE (& PENCILS)

MDR report key: 556200 · Received November 19, 2004

Report

Report Number
556200
Event Type
Other
Date Received
November 19, 2004
Date of Event
November 12, 2004
Report Date
November 19, 2004
Manufacturer
VALLEYLAB - TYCO HEALTHCARE GRP US-SURGICAL
Product Code
GEI
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

BOVIE BURN MACHINE WAS THREADED THROUGH DRAPES DURING OPEN HEART SURGERY. WHEN TOWEL CLIPS WERE ATTACHED TO THE DRAPES, IT CUT THROUGH BOVIE WIRE WHICH WAS INADVERTENTLY LYING ON PT'S SKIN. STAFF UNAWARE UNTIL AFTER SURGERY. BURN APPROXIMATELY 3 INCHES LONG. SURGEON IMMEDIATELY ADDRESSED BY SUTURING AREA. AREA CLEANED AND SUTURED AT TIME OF OCCURRENCE. PT AWARE OF INCIDENT. PT DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BOVIE BURN MACHINE (& PENCILS) SAME GEI VALLEYLAB - TYCO HEALTHCARE GRP US-SURGICAL T3J29121A A FORCE FC.C. *

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other