FDA Adverse Event Malfunction Summary report: N

CARELINK ENCORE

MDR report key: 5559259 · Received April 7, 2016

Report

Report Number
3004593495-2016-00133
Event Type
Malfunction
Date Received
April 7, 2016
Date of Event
February 17, 2016
Report Date
February 17, 2016
Manufacturer
PLEXUS MANUFACTURING SDN. BHD
Product Code
KRG
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; THE PROGRAMMER POWERED UP WITH A SYSTEM ERROR. ANALYSIS ALSO FOUND THE PROGRAMMER HAD A FFT (FAST FOURIER TRANSFORM) FIRMWARE ERROR WHEN RUNNING FFT SOFTWARE WHILE RF (RADIO FREQUENCY) HEAD WAS PLUGGED INTO THE PROGRAMMER DUE TO THE RF HEAD DIGITAL PRINTED CIRCUIT BOARD ASSEMBLY. (B)(4).

Additional Manufacturer Narrative · 1

FOLLOWING SERVICE AND REPAIR OF THE PROGRAMMER, FAILURE ANALYSIS WAS PERFORMED ON THE RF HEAD DIGITAL PRINTED CIRCUIT BOARD ASSEMBLY (PCBA). VISUAL INSPECTION REVEALED NO ANOMALIES. THE PCBA WAS ASSEMBLED INTO A KNOWN GOOD UNIT. THE PROGRAMMER WAS THEN POWERED UP. THERE WAS AN ERROR UPON BOOT. AN AUDIBLE HIGH-PITCHED NOISE WAS NOTED COMING FROM THE RF HEAD. THERMAL IMAGING INDICATED TWO ABNORMALLY HEATED COMPONENTS. VERIFIED NO 12-VOLT OUTPUT, THE RF HEAD DIGITAL PCBA WAS DEFECTIVE. WHEN THE PCBA WAS REPLACED, PROPER OPERATION RESULTED. THE SUSPECT PCBA WAS REINSTALLED AND ABNORMAL OPERATION RETURNED. CONCLUSION: THE EVENT SEEN DURING SERVICE AND REPAIR WAS CONFIRMED, THE RF HEAD DIGITAL PCBA WAS DEFECTIVE.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY :ANALYSIS CONFIRMED THE REPORTED EVENT; THE PROGRAMMER POWERED UP WITH A SYSTEM ERROR. ANALYSIS ALSO FOUND THE PROGRAMMER HAD A FFT (FINAL FUNCTIONAL TESTING) FIRMWARE ERROR WHEN RUNNING FFT SOFTWARE WHILE RF (RADIO FREQUENCY) HEAD WAS PLUGGED INTO THE PROGRAMMER DUE TO THE RF HEAD DIGITAL PRINTED CIRCUIT BOARD ASSEMBLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER DISPLAYED AN ERROR UPON BOOT-UP, WHICH WAS NOT ABLE TO BE CLEARED BY RE-BOOTING. THE PROGRAMMER HAS BEEN RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212351 CARELINK ENCORE PROGRAMMER, PACEMAKER KRG PLEXUS MANUFACTURING SDN. BHD 29901

Patients

Seq Age Sex Outcome Treatment
1